Rear Panel Functions
Graphic
Description
Equipotential Ground Stud
– Provides a means of
securely linking the earth grounds of the
AtriCure ASU to other grounded
equipment.
Data Port
– For manufacturing and test purposes.
Power Entry Module
– This module contains both
the ON/OFF switch and the fuses. The
voltage is selected by the orientation of
the fuse drawer as marked.
Fuse Box
– The Fuse Box contains fuses selected for
the input voltage. See Technical
Specifications in Section 7 of this manual.
Input Voltage Selector Switch
– The input voltage
selector switch is pre-set at the factory to
either 110V or 220V and
should not
be
adjusted by the operator. This setting
should only be adjusted by the
manufacturer or by an authorized service
representative.
Speaker Volume Control
–
The audible volume
level is adjustable via a rotary dial.
The ASU includes a speaker for
producing audible feedback to the user.
Footswitch Receptacle
– This receptacle accepts
the Footswitch connector. The single
momentary actuation pedal provides for
the activation of RF power output.
Service Access
– For manufacturing and test
purposes.