USER’S MANUAL
Page 16 of 39
P000463.F
3.
Installing the ASU
Inspect the ASU for any signs of physical damage to the front panel, chassis or cover.
NOTE:
If any physical damage is found, DO NOT USE THE UNIT. CONTACT AtriCure
for a replacement
.
All returns must be approved by AtriCure.
3.1.
Transporting the ASU
The handle may be used to carry the ASU. To change the positioning of the handle, depress both
handle adjustment knobs simultaneously and move the handle to the desired location.
Do
not
change the handle position when a Handpiece or Accessory Device is connected to the Handpiece
receptacle.
3.2.
Adjusting the Viewing Angle
To change the viewing angle of the ASU Conductance Graph Display, adjust the handle position
using the directions in Section 3.1., above.
3.3.
Preparing the ASU For Use
The ASU may be placed on a mounting cart or on any sturdy table or platform. It is recommended
that carts have conductive wheels. Refer to hospital procedures or local codes for detailed
information.
Provide at least four to six inches of space around the sides and top of the ASU for convection
cooling. Under continuous use for extended periods of time, it is normal for the top and rear panel
to be warm.
3.4.
Power Cord
The ASU is shipped with an approved hospital grade power cord.
Plug the ASU into a grounded receptacle.
NOTE:
Do not use extension cords or three-prong to two-prong adapters. The Power Cord
assembly should be periodically checked for damaged insulation or connectors.