APD series bench scale   

 

 

 

 

 

 

    

 

 

 

 

 

 

 

 

 

 

 

 

          APD_01_11.05_EN_U 

 

 

5. POWER SUPPLY AND START UP 

 

5.1 POWER SUPPLY 

The instrument can be powered with a 6 Vdc tension supplied from an AC/DC external adapter which should be connected 
to the 110-240 Vac mains voltage. Safety norms must be respected for the connection to the mains voltage including the 
use of a line which has to be free from noise generated by other electronic equipment.  
It is also possible to power the scale through the rechargeable built-in battery. 

 

6V RECHARGEABLE BATTERY FEATURES 

Material   Lead   
Power   4,5 Ah   
Output  6 V   

THE BATTERY MUST ONLY BE REPLACED WITH AN ORIGINAL FROM THE MANUFACTURER

 

In order 

TO POWER 

the instrument through the 110-240Vac mains or 

TO RECHARGE

 the 6V battery, insert the plug end 

of the AC/DC power adapter into the socket in the lower part of the indicator (chapter “INSTALLATION”) and the adapter in 
your main current source.  

 

Do not connect other equipment to the same socket as the one that the adapter is in. 
Do not step on or crush the power supply cable. 
 

 

5.2 TURNING ON / OFF THE INSTRUMENT 

 

TO TURN ON 

the scale press the 

MODE/C

 key until the indicator turns on; then release. 

The display shows: 

XX.YY

 

 is the installed software version. 

The instrument turns on all the display segments and symbols.  

MAX XXX.XXX

               capacity of channel 1  

bt  XXX

                           in which XXX is a number from 0 to 100 which indicates the battery level (if present). 

 
The scale has an “auto zero at start-up” function: in other words it means that if at start-up a weight /- 10% of the 
capacity is detected, it will be zeroed; if the weight is not within this tolerance, with a non approved instrument the display 
shows  the  present  weight  after  a  few  instants,  while  with  an  approved  instrument  “ZEro”  is  shown  continuously  on  the 
display, until the weight does not re-enter within this tolerance; the auto zero function at start-up may be disabled in the set-
up  environment  (only  with  non  approved  instrument);  see 

SEtuP 

>>

  ConFiG 

>>

  PArAM 

>>

  Auto-0 

parameter

 

(TECH.MAN.REF.)

 

By pressing the 

ZERO

 

key for an instant while the version is shown in the display, the indicator will show the following in 

this order: 

XX.YY 

 

in which XX indicates the instrument type, YY indicates the metrological software version. 

XX.YY.ZZ

 

  

is the installed software version. 

APd          

 

 

 

 

is the name of the installed software. 

bt  XXX

 

 in which XXX is a number from 0 to 100 which indicates the battery level (if present). 

-K- X.YY

  

 in which K identifies the type of keyboard: K=0  5-key keyboard. 

 

X.YY is the installed software version. 

 
After this, it shows the programmed capacity and the minimum division, “hi rES” (in case of NOT approved instrument) or 
“LEGAL” (in case of approved instrument), the g gravity value, and finally it executes a countdown (self-check). 

 
TO RESTART 

the scale keep the 

MODE/C

 key pressed for 8 seconds long: the indicator will turn off, then release the key 

and the indicator will automatically turn on.  

 
TO TURN OFF 

the scale keep the 

MODE/C

 key pressed until the “- OFF–“ message appears on  the display; then release 

the key. 

 

Summary of Contents for APD Series

Page 1: ...an be mounted attached to the platform or can be mobile APD_01_11 05_EN_U User Manual ATP Messtechnik GmbH J B von Weiss Strasse 1 D 77955 Ettenheim Email info atp messtechnik de Internet www atp messtechnik de Tel 0 7822 8624 0 FAX 0 7822 8624 40 ...

Page 2: ...8 2 TARE OPERATIONS 11 8 3 LIMITATION OF THE TARE FUNCTIONS FOR DIRECT SALE 12 8 4 AUTO POWER OFF FUNCTION 12 8 5 LOW BATTERY WARNING 12 8 6 MULTI RANGE FUNCTIONING for legal for trade approved instruments 13 8 7 CONVERSION OF THE UNIT OF MEASURE 13 8 8 DISPLAY OF METRIC DATA inFO 13 9 TOTALISATION 14 9 1 OPERATIONS OF TOTALISATION 14 9 2 REACTIVATION OF THE TOTALISATION 15 9 3 VISUALIZATION OF TH...

Page 3: ...y has to be completely recharged 12 hours in the first installation of the instrument we RECOMMEND disconnecting the battery if the instrument is not going to be used for more than 30 days In order to avoid the deterioration of the rechargeable battery In standard conditions never leave the battery partially or completely uncharged at least once a week recharge it completely In case the instrument...

Page 4: ...ainless steel WEIGHING PLATE TARE FUNCTION Available on the entire capacity AUTO POWER OFF Programmable from 1 to 255 minutes or disinserted LOW BATTERY WARNING Low Batt will appear on the display BATTERY RECHARGE TIME 12 hours LOAD CELL POWER SUPPLY 5Vdc 5 120mA max 8 cells of 350 Ohms LOAD CELL CONNECTIONS 6 wires CELL1 with Remote Sense 4 wires CELL 2 3 4 without Remote Sense THE PARTS OF THE I...

Page 5: ...APD series bench scale APD_01_11 05_EN_U 5 4 INSTALLATION 1 Input for the external power supply 2 Input for the cable of the cell and battery power supply 1 2 2 ...

Page 6: ...ght indicator has an ABS case whose external dimensions are represented in figures The indicator can be mounted attached to the platform or it can be installed in any desired position In the following figures the dimensions of the scale and the plate are shown ...

Page 7: ...cale has an auto zero at start up function in other words it means that if at start up a weight within 10 of the capacity is detected it will be zeroed if the weight is not within this tolerance with a non approved instrument the display shows the present weight after a few instants while with an approved instrument ZEro is shown continuously on the display until the weight does not re enter withi...

Page 8: ...f the scale is unloaded In the numeric input phase it increases the digit to be modified M It carries out the totalization if the conditions for this operation are satisfied In the numeric input phase it selects the digit to be modified from left to right MR If pressed for an instant it carries out the display of the number of totalizations and the present total In the numeric input phase it confi...

Page 9: ...ge is selected 6 B The displayed value is a gross weight if the German French or Spanish language is selected 7 Indicates the battery charge level see the section LOW BATTERY WARNING 8 MAX MIN e When viewing the metric information it identifies the indicated capacity range When viewing the metric information it identifies the indicated minimum weigh range When viewing the metric information it ide...

Page 10: ...easure in use kg for kilogram g gram 15 Not used 16 t Indicates the unit of measure in use tons 17 LB Indicates the unit of measure in use pounds 18 Not managed 19 Indicates that a key has been pressed 20 PEAK Not used 21 HOLD Not used 22 SP1 SP2 SP3 SP4 Not used Not used Not used Not used ...

Page 11: ...the value will be subtracted from the weight present on the plate and the relative pilot lights will turn on NOTE If the entered value is not a multiple of the scale s minimum division it will be rounded off CANCELLING A TARE One can manually cancel the tare value in different ways unload the scale and press the TARE or ZERO key carry out the tares in deduction partially unloading the scale and pr...

Page 12: ...imes to scroll ahead through the parameters or TARE to scroll backwards until one finds the AutoFF parameter Press MR to enter the parameter With the ZERO or TARE keys select the possible options diSAb auto switch off disabled EnAb auto switch off enabled Confirm with MR if EnAb has been selected one will be asked to enter the number of minutes after which the scale should turn off enter a number ...

Page 13: ...ion follow the procedure below Turn on the scale press the TARE key while the firmware version is displayed the display shows the F ModE step Press MR to enter the step Press ZERO many times to scroll forwards through the parameters or TARE to scroll backwards until one finds the un ConV step Press MR to enter the parameter With the ZERO or TARE keys select the possible options diSAbL disables the...

Page 14: ...mber of totalisations set in this step the accumulated total is shown on the display and cleared one can set a value between 0 and 255 NOTE the value 0 disables the function 9 1 OPERATIONS OF TOTALISATION This scale allows to totalize the weight in load and unload Totalisation in load If the manual totalisation has been set in order to carry out the totalisation it is necessary to load the weight ...

Page 15: ...stic signal is emitted NOTES If the gross or net weight in absolute value is less than the requested minimum weight by pressing the M key the display shows the LoW error message If the scale is in the under load or over load status by pressing the M key the display shows the un oVEr error message 9 3 VISUALIZATION OF THE TOTAL By pressing the MR key one can temporarily view on the display the numb...

Page 16: ...tP2 o tP3 is equal to the zero point tP0 Er 37 The number of converter points per scale division is less than two Carry out again the calibration with special attention to the capacity and the division Er 39 It is displayed when the instrument has not yet been calibrated and initialized Press the TARE key when the instrument displays Er 39 to enter the technical set up environment Carry out the in...

Page 17: ...ary for any reason where the instrument is used the Customer will pay for all of the service technician s costs travel time and expenses plus room and board if any The Customer pays for the transport costs both ways if the instrument is shipped to DEALER or manufacturer for repair The WARRANTY is VOIDED if any of the following occurs repairs or attempted repairs are made by unauthorised personnel ...

Reviews: