18
O&M Manual
Rev-A (6/19)
5
Status Flags are as follows -
Bit
Status 1
Status 2
0 (LSB)
ALARM_A
EE_INIT_FAIL
1
ALARM_B
MAIN_UNITS_HI
2
ALARM_C
MAIN_UNITS_LO
3
ALARM_D
MAIN_INPUT_ERR
4
ALARM_E
TC_UNITS_HI
5
ENTRY_OUT_OF_R
ANGE
TC_UNITS_LO
6
ENTRY_ACCEPTED
TC_INPUT_ERR
7
ENTRY_FAIL
CAL_MAIN_SLOPE_HI
8
MAIN_CAL_PASS
CAL_MAIN_SLOPE_LO
9
MAIN_CAL_FAIL
CAL_MAIN_ZERO_HI
10
TC_CAL_PASS
CAL_MAIN_OFFSET_H
I
11
TC_CAL_FAIL
MAIN_UNSTABLE
12
TC_F
CAL_TC_OFFSET_HI
13
SENSOR_LOCK
TC_UNSTABLE
14
NU
NU
15
NU
NU
Sensor 1 INFO
Bytes
Data Type
Description
212-213
INT
S1 Slope
214-215
INT
S1 Offset
216–217
INT
S1 Delay
218–219
INT
S1 Alarm A
220-221
INT
S1 Alarm B
222-223
INT
S1 Slope Alarm
224-225
INT
S1 Timer Limit
226-227
INT
S1 Vout HI
228-229
INT
S1 Vout LO
230-231
INT
S1 TC Mode
232-239
LWORD
S1 Tag1
240-247
LWORD
S1 Tag2
Sensor 2 INFO
Bytes
Data Type
Description
248-249
INT
S2 Slope
250-251
INT
S2 Offset
252–253
INT
S2 Delay
254–255
INT
S2 Alarm A
256-257
INT
S2 Alarm B
258-259
INT
S2 Slope Alarm
260-261
INT
S2 Timer Limit
262-263
INT
S2 Vout HI
264-265
INT
S2 Vout LO
266-267
INT
S2 TC Mode
268-275
LWORD
S2 Tag1
276-283
LWORD
S2 Tag2