Manual, F/T Sensor, Ethernet Axia
Document #9610-05-Ethernet Axia-09
Pinnacle Park • 1031 Goodworth Drive • Apex, NC 27539 • Tel:+1 919.772.0115 • Fax:+1 919.772.8259 •
46
6.2 Snapshot Page (rundata.htm)
This page displays the current sensor loading conditions. The information on the
Snapshot
page is static and
becomes current after a user loads or reloads the page.
Figure 6.3—Snapshot Page
Transducer Loading Snapshot (User Units):
Force/Torque Data:
These fields display the force and torque data scaled in the user units that the user
selected on the
Configurations
page. If any strain gages are overloaded, these values
are invalid and displayed in red with a line through them.
Transducer Loading Snapshot (Counts):
Force/Torque Counts:
These fields display the force and torque data scaled with the counts per force
and counts per torque displayed on the
Configurations
page. For more information
about how F/T values are scaled, refer to
Section 4.4—Converting Counts Per Force/
. If any strain gages are overloaded, these values are invalid and
displayed in red with a line through them.
Strain Gage Data:
Unbiased Gage Data:
These fields display the sensor’s raw strain gage information for troubleshooting
overload errors. If the strain gages are overloaded, the values are invalid and
displayed in red.
NOTICE:
• When an overload condition occurs, the reported force and torque values are invalid.
• Individual strain gage values do not correspond to individual force and torque axes.
• The sensor readings on this page are captured as the web page requests them. It is
possible that the readings towards the bottom of the page have come from later F/T data
records than the readings towards the top of the page.