Chapter 3 –
Start up, adjust and measures
UM-24100D-U User
manual
ATEQ CDP60
Page 12/36
3. PARAMETERS ADJUST
It's existing three adjusted parameters.
3.1. F
ILTER
It can be adjusted from 0.1 to 9.9 seconds by 0.1 second of step. It allows making a
mean on the measurement time, making easier the measurement reading.
Enter into the parameters menu, by pressing and
old the
ENTER
during 3 seconds.
500.0 Pa
000.0
The
PARAMETER
menu is displayed. By using the
UP
and
DOWN
key select
FILTER
parameter.
PARAMETER
>FILTER : 0.5 s
PRESS.UNIT.: Pa
BACK LIGHT: 100 %
Then press the
ENTER
key to modify the
parameter (the cursor moves on the right).
PARAMETER
FILTER : 0.5 s <
PRESS.UNIT.: Pa
BACK LIGHT: 100 %
By using the
UP
and
DOWN
key
modify the parameter to the hoped value.
PARAMETER
FILTER : 1.0 s <
PRESS.UNIT.: Pa
BACK LIGHT: 100 %
Validate with the
ENTER
key.
PARAMETER
>FILTER : 1.0 s
PRESS.UNIT.: Pa
BACK LIGHT: 100 %
To return to the measurement mode, press
CANCEL
.
500.0 Pa
000.0