PXIe-1209 User Manual
Publication No. 981050 REV. A
Operation 5-12
Astronics Test Systems
The next step is to add the various driver VIs that provide the information you need
in your application. The image below demonstrates that three more VIs have been
added to the diagram. These VIs set the pulse repetition frequency, pulse width,
and pulse amplitude for the selecte pulse generator (PGEN 1).
One requirement for using the driver is to wire the “instrument handle out” output
from one VI to the “instrument handle” input on the next VI. A second requirement
for using the driver is to wire the “error out” output from one VI to the “error in”
input of the next VI. These two required connections are shown for each of the 3
newly added VIs:
The last recommendation, as with the LabWindows/CVI driver, is that every handle
you open with the “ri1209e Initialize.vi” should be closed with the “ri1209e
Close.vi”. Every time you execute the “ri1209e Initialize.vi”, it opens a new handle
Summary of Contents for PXIe-1209
Page 22: ......