Error processing
LDS70A User Manual
Page 56
ASTECH GmbH
6
Error processing
In case of one or more detected errors, the error code with the highest value
will be transmitted (most serious defect/error).
Example:
no distance value identified, no other error: DE02
no distance value identified, in addition temperature out of range:
DE06
Non-critical errors will be reset automatically, a critical error requires a device
reset (see table)
–
command DR or power off/ on.
Hardware errors generate error code DE04.
Table 13: Error codes
Error code
Description
Reset
DE02
No distance identified
Automatically, with the next
measured distance
DE04
Device error (hardware)
Not automatically, device reset
necessary
DE06
Temperature out of range
Automatically, if temperature
returned to specified range
DE10
Internal
laser
voltage
lower than the defined
minimum
voltage
(ULaser
MIN
)
Not automatically, device reset
necessary
In binary format all errors will be sent as “0” (00). The described error codes
will be outputted, if an error is detected during
- LDS70A measurement (DM, DT)
- a measuring mode (DM, DT) will be started.
Output of error message (like output of measuring values):
- once for measuring mode DM
- in the specified output frequency for measuring mode DT