LDS70A User Manual
Parameter setup and measurement
ASTECH GmbH
Page 29
5
Parameter setup and measurement
5.1
General information
The LDS70A is parameterized using the serial interface. Precondition for
programming via UART is a connection provided by a terminal program (e.g.
HyperTerminal, Tera Term).
5.2
Transmission protocol
Both serial interfaces use the same transmission protocol:
▪
Interface settings: Asynchronous, 8 data bits, no parity, 1 stop bit
▪
Transmission protocol format / syntax: 7 bit ASCII
▪
Proprietary transmission protocol
▪
Commands are case-insensitive (no differentiation between lower and
upper case).
▪
Decimal separator in the output of figures is the dot “.” (0x2E).
▪
The terminator of a command (sending command) is ENTER (0x0D, 0x0A)
or Carriage Return (0x0D) or Line Feed (0x0A)
▪
Where parameters have several values, they are separated by a space
(0x20).
▪
The response to commands with parameters is the respective command
including the parameters.
▪
The response to commands without parameters is the respective
command including the current parameters.
▪
The response to commands with parameters outside of the valid value
range is the respective command including the current parameters.
▪
The response to unknown commands and faulty parameter formats is a
“?” (0x3F).