Parameter setup and measurement
LDS70A User Manual
Page 52
ASTECH GmbH
Trigger setup
–
general information
The LDS70A Trigger may be used as input or output. There are three possible
scenarios:
1. Trigger input
–
externally initiated measurement
External trigger signal will be sent to LDS70A to start a single
measurement like DM in accordance with parameter TI.
2. Trigger output - connection between two devices:
The output trigger signal of the 1. LDS70A (parameterized with TO)
starts a single measurement DM of the second LDS70A
(parameterized with TO).
3. Continuous distance tracking controlled by trigger
Differences between trigger input and trigger output:
Important is the parameter y of TI and TO.
TI y > 0 / TO y = 0
Trigger input
The measurement starts after an external trigger
impulse.
TI y = 0 / TO y >0
Trigger output
LDS70A sends a trigger impulse to the second
device.
The parametrization of the trigger connection is carried out via the serial
interface.
For the trigger function either TI
or
TO must be activated only.
A concurrent use of TI and TO is not possible.