LDS70A User Manual
Device description
ASTECH GmbH
Page 23
4.8
Serial interface RS422
The RS422 serial interface can be used for data transfer and configuration of
the LDS70A like the RS232. Cable length depends on the baud rate and the line
quality; up to 300 m length is possible. Use only high-quality cable with 2
twisted pairs and shielding.
Pin 3 (white) of the connector decides which kind of serial interface active. For
RS422 the pin has to be connected to GND (pin 12, red/blue).
VDC +
VDC -
TX+
RX+
RS422
brown
blue
white
green
pink
yellow
black
grey
red
violet
grey/pink
red/blue
shield
1
2
3
4
5
6
7
8
9
10
11
12
sh
power supply
10
V …
30 V DC
-
+
RS422
RX-
TX-
RS232/RS422
Figure 4: Wiring with serial interface RS422