Ramp-DOWN
: A length of time that is allowed for the test voltage to decay from set
point to zero.
Arc Detect
: This function when turned on activates the Arc Detection threshold
controlled by the Arc Sense parameter.
Arc Sense
: The numbers corresponding to the different arc sensitivity levels are 1
through 9, 1 meaning the maximum threshold of allowable arcing, 9 meaning the
minimum threshold of allowable arcing. During Hipot testing, some low current arcing
may be allowable. Arc sense is a maximum allowable threshold for arcing.
Frequency
: This parameter is available in AC tests only and is selectable using the
settings menu between 50 and 60Hz.
Connect
: This function when turned ON will connect or link the current memory to the
next memory. This function allows all 50-memory locations to be connected together.
Ramp-HI
: Ramp-HI will allow current higher than the normal HI-Limit current setting of
the DC Withstand Voltage test, during the Ramp-Up time, to avoid false failure due to
charging current. Ramp-HI may be selected ON or OFF using a soft key. A
description of how to set up this parameter is given in the
4.6.10. Ramp-Hi selection
section of this manual.
Charge–LO
: The Charge-LO function is used to check if the cables are connected
properly at the beginning of a test. This function is only available in DC Withstand and
Insulation resistance testing. A description of how to set up this parameter is given in
the
4.6.11. Charge-Lo setting