4.5.3. Default Test Parameters
The following table is a listing of the Default Parameters for the ACW tests available in
the 7705 and 7715 HypotMAX and the DCW tests available in the 7710 and 7720
HypotMAX When you receive your HypotMAX all 50 memory locations will be loaded
with these parameters. The Default Parameters are as follows:
TEST TYPE
PARAMETER
VALUE
ACW units
Voltage
1.24kVAC
HI-Limit (current)
5.000mA
LO-Limit (current)
0.000mA
Ramp UP
0.3s
Dwell Unit
Second
Dwell Time
1.0s
Ramp-DOWN
0.0s
Frequency
60Hz
Arc Detect
OFF
Arc Sense
5
Connect
OFF
DCW
units Voltage
1.50kVDC
HI-Limit
5000
µ
A
LO-Limit
0.0
µ
A
Ramp-UP
0.4s
Dwell Unit
Second
Dwell Time
1.0s
Ramp-DOWN
0.0s
Arc Detect
OFF
Arc Sense
5
Ramp-HI
OFF
Charge-LO
0.0
µ
A
Connect
OFF
4.6. Setting Up a Test
Before going to setup the Test Parameters, make sure that the Keyboard is in the
Unlock mode, and then follow this procedure to setup the Test Parameters.
Press the Memory key and using the numeric keypad enter the Memory number you
would like to edit and then press the ENTER key.
Use the
∧
or
∨
arrow keys to progress through the test parameters menu. The
∨
key
will advance forward and
∧
key will advance backward. The sequential forward menu
items are Voltage, HI-Limit, LO-Limit, Ramp-UP, Dwell Unit, Dwell Time, Ramp-
DOWN, Frequency, Arc Detect, Arc Sense, and Connect.