16
Ramp-
DOWN
Range:
0.0 – 999.9 sec
Resolution:
0.1 sec increments
Accuracy:
±
(0.1% + 1 count)
Ground Fault GFI Trip Current:
1 mA max
Interrupt
HV Shut Down
response:
< 1 ms
3.2. 7710 Functional Specifications
INPUT
Voltage
115/230 VAC
±
10
%
, Single Phase, User selection
Frequency
50/60 Hz
±
5
%
Fuse
6.3 Amp, 250V Slow Blow
DIELECTRIC WITHSTAND TEST MODE
Output Rating
12 kV @ 10 mA
Output Adjustment
Range:
0.00 – 12.00 kV DC
Resolution:
10 volt/step
Accuracy:
±
(2
%
of setting
+
10 volts)
Ramp-HI
13 mA peak maximum, 10 mA DC, ON/OFF selectable
Charge-LO
Range:
0.0 - 350.0
µ
A DC or Auto set
HI and LO-Limit
Range 1:
0.0 – 999.9
µ
A
Resolution:
0.1
µ
A/step
Range 2:
1000 – 9999
µ
A
Resolution:
1
µ
A/step
Accuracy:
±
(2
%
of setting
+
2 counts)
Arc Detection
Range:
1 – 9
Failure Detector
Audible and Visual
Voltage Display
Range:
0.00 – 12.00 kV Full Scale
Resolution:
10 volt
Accuracy:
±
(2
%
of reading
+
20 V)
Current Display
Auto Range
Range 1:
0.0
µ
A - 350.0
µ
A
Resolution:
0.1
µ
A
Range 2:
300
µ
A - 3500
µ
A
Resolution:
1
µ
A
Range 3:
3000
µ
A - 9999
µ
A
Resolution:
10
µ
A
Accuracy:
±
(2
%
of reading
+
3 counts)
DC Output Ripple
≤
5
%
Ripple RMS at 12 kV DC @ 9999 uA, Resistive
Load