35
4.1.4.
Step Selection
Press the STEP key, the display will show the test parameters of the functions selected
for that step, i.e. the parameters of AC Withstand Voltage , DC Withstand Voltage,
Insulation Resistance or Ground Bond test. Pressing the step key again will advance to
the next step in sequence with step 1 following step 8.
Note:
When the steps have been interconnected you will see an underbar character on
the display just after the step number. This indicates that the step after the active step
will run immediately after the active step is completed. The display will show:
ACW Set XXX.X s
M25-1_ X.XX KV XX.XX mA
4.2.
Setting Up Test Parameters
Before going to setup the Test Parameters, make sure that the Keyboard is in the Unlock
mode, then follow this procedure to setup the Test Parameters.
4.2.1.
AC Withstand Test Parameter Setting
Press the ACW key and the display will show:
ACW Set
XXX.X s
MXX-X X.XX KV XX.XX mA
Note: X = Numeric
ACW Set
: AC Withstand Voltage test setting screen
XXX.X s
: Dwell Time setting
MXX
: Memory Program number
- X
: Test Step number
X.XX KV
: AC Output Voltage setting
XX.XX mA : AC High-Limit current setting
Use the
or
arrow keys to progress through the test parameters menu. The
key will
advance forward and
key will advance backward. The sequential forward menu items
are Voltage, HI-Limit, LO-Limit, Ramp Time, Dwell Time, Frequency, Arc Sense, Arc
Fail, Scanner Set, Offset, and Connect.
AC Output Voltage setting
Advance the menu to the Voltage parameter. The display will show:
Voltage = XXXX V
Range : 0 - 5000