61
If the DUT current is well beyond the metering range of AC Withstand Voltage test and
an Arcing condition beyond the Arc Sense limit is indicated, the display will show:
ACW Breakdown
XXX.X s
MXX-X X.XX KV
40 mA
If the DUT current falls below the LO-Limit of AC Withstand Voltage test the display
will show:
ACW LO-limit
XXX.X s
MXX-X X.XX KV XX.XX mA
If the DUT current is within the metering range of the AC Withstand Voltage test and an
Arcing current exceeds the Arc-Sense limit and the Arc function is set to ON, then an Arc
failure has occurred and the display will show:
ACW Arc-Fail
XXX.X s
MXX-X X.XX KV XX.XX mA
If the GFI threshold is exceeded during of the AC Withstand Voltage test, the display will
show:
ACW GND-Fault XXX.X s
MXX-X X.XX KV XX.XX mA
When the DUT passed the AC Withstand Voltage test, when the test process is complete
the display will show:
ACW Pass
XXX.X s
MXX-X X.XX KV XX.XX mA
5.7.3.
DC Withstand Displayed Messages
If the test in process is Aborted with the RESET button or remote control, the display will
show:
DCW Abort
XXX.X s
MXX-X X.XX KV XXXX
A