39
Use the ENTER key to select the Step Connection mode, then press
or
key to
advance the program to another test parameter or press the EXIT key to exit from the
setting mode to the operation mode.
If Connect is set to ON, the next step in the sequence will be executed. If Connect is set
to OFF, the test sequence will stop at this step.
If Step 8 is set to ON the test process will be connected to the first step of the next
Memory.
4.2.2.
DC Withstand Test Parameter Setting
Press the DCW key and the display will show:
DCW Set XXX.X s
MXX-X X.XX KV XXXXX
A
Note: X = Numeric
DCW Set
: DC Withstand Voltage test Setting screen
XXX.X s
: Dwell Time setting
MXX
: Memory Program number
- X
: Test Step number
X.XX KV
: DC Output Voltage setting
XXXX
A : DC High-Limit current setting
Use the
or
arrow keys to progress through the test parameters menu. The
key will
advance forward and
key will advance backward. The sequential forward menu items
are Voltage, HI-Limit, LO-Limit, Ramp Time, Dwell Time, Charge-LO, Ramp-HI, Arc
Sense, Arc Fail, Scanner Set, Offset, Connect.
DC Output Voltage setting
Advance the menu to the Voltage parameter. The display will show:
Voltage = XXXX V
Range : 0 – 6000
Use the Numeric keys to enter the voltage setting, then press the ENTER key. The
program will store the voltage setting and advance to the HI-Limit parameter
automatically. The unit is volt and 1 volt per step.
Press the EXIT key to exit from the setting mode to the operation mode if all parameters
have been set.