55
5.
Operating Instructions
5.1.
Setup
Before the operation of this instrument, make sure that all Test Parameters have been set
properly according to the Test Parameters Setup Procedures. Also check the system
setting of Remote Control, LCD Contrast , the Alarm Volume and Fail Stop.
Be sure to connect the appropriate test leads to the device under test (DUT) or test
fixture. Be sure to connect the safety ground (on the rear panel) to a suitable known good
ground before energizing this instrument. Then connect the return lead first to the test
fixture or the DUT followed by the high voltage lead.
Check your connections to be sure they are making good contact and that the test station
or area is clear of debris or other personnel.
DO NOT TOUCH THE DEVICE UNDER TEST ONCE
THE TEST HAS BEEN STARTED.
5.2.
Power-Up
Turn on the Input Power Switch. The display will show the Trade Mark, Model Number
and Version Number first, as follows:
ARI HYPOTMAX
7704 VER : X.X
Note: X = Numeric.
Then, all status LEDs will light up, and the Scanner status LEDs will light sequentially if
the internal scanner has been installed. The program will then recall the Memory and
Step which was last executed.
5.3.
Settings Screen
The settings screen will show parameter settings of the test that will be performed and the
memory locations that they are stored in. This screen can be accessed after a test is
Aborted or Passed by pressing the RESET button, or after a test failure by pressing the
RESET button twice. The Function is indicated on the left side of the first line and the
mode, or test result, is indicated just to the right of the Function. The timer is always
shown on the right side of the first line. The left side of the second line always indicates
the Memory and Step that has been selected and an underbar character to the right of the
step number indicates that the step has been connected to the next step in sequence. The
display will show the underbar as follows for an ACW test, and step 1 connected to step
2:
ACW Set XXX.X s
M25-1_ X.XX KV XX.XX mA