64
When the DUT passed the DC Withstand Voltage test, when the test process is complete
the display will show:
DCW Pass
XXX.X s
MXX-X X.XX KV XXXX
A
5.7.4.
Insulation Resistance Displayed Messages
If the test in process is Aborted with the RESET button or remote control, the display will
show:
IR Abort
XXX.X s
MXX-X XXXX V XXXX M
If the test in process is Aborted with the RESET button or remote control before the
meter readings are taken, the display will show:
IR Abort
XXX.X s
MXX-X - - - - V - - - - M
or
IR Abort
XXX.X s
MXX-X XXXX V - - - - M
At the beginning of Insulation Resistance test when the voltage begins to ramp but before
the meter readings are taken, the display will show:
IR Delay
XXX.X s
MXX-X - - - - V - - - - M
During the Insulation Resistance test when the values are being updated in real time
during the Delay cycle, the display will show:
IR Delay
XXX.X s
MXX-X XXXX V XXXX M
If the leakage current during the Ramp cycle falls below the Charge -LO setting, then the
display will show:
IR Charge-LO
XXX.X s
MXX-X - - - - V - - - - M
If the DUT current exceeds the HI-Limit of Insulation Resistance test and the resistance
is within the metering range, the display will show:
IR HI-Limit
XXX.X s
MXX-X XXXX V XXXX M