Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The
ASSKEA pro
wound
®
M and ASSKEA pro
wound
®
S
devices are intended for operation in the
electromagnetic environment specified below.
The customer or user of the
ASSKEA pro
wound
®
M or ASSKEA pro
wound
®
S
must ensure that it
is used in such an environment.
Immunity tests
IEC 60601 test
level
Compliance level
Electromagnetic Environment
– Guidance
Conducted RF
disturbances
pursuant to
IEC 61000-4-6
Radiated RF
disturbances
pursuant to
IEC 61000-4-3
3 V
eff
150 kHz to
80 MHz
3 V/m
80 MHz to
2.5 GHz
3 V
eff
3 V/m
Portable and mobile RF
communications equipment
should be used no closer to
any part of the
ASSKEA pro
wound
®
M and
ASSKEA pro
wound
®
S
, incl.
cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance:
d= 1.2 √P
d= 1.2 √P for 80 MHz to 800
MHz
d= 2.3 √P for 800 MHz to 2.5
GHz
Where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in meters (m).
Field strength from fixed RF
transmitters, as determined
by an electromagnetic site
survey
a
, should be lower than
the compliance level in each
frequency range.
b
Interference is possible in the
vicinity of devices bearing the
symbol shown below.