3.3
Patient Mode
The
ASSKEA pro
wound
®
M and ASSKEA pro
wound
®
S
enable the selection of the patient mode
at startup. In patient mode, the settings for the alarm sensitivities “System closed” and “Check
dressing seal” as well as the patient runtime can be viewed and reset. To select the patient mode,
follow these steps:
1.
Press the
button for 1-2 seconds to switch on the
ASSKEA pro
wound
®
M or
ASSKEA pro
wound
®
S
. The following start screen is displayed for 5 seconds:
2.
While the start screen is displayed, press and hold down the
button and additionally
press the
button for 1-2 seconds. The
Authorization
screen is displayed.
3.
Use the
arrow buttons to enter the “xxxx” code.
Press the
arrow button until the desired digit of the code is displayed and confirm the
entry with the
button. Select the other digits of the code with the
arrow button and
confirm them with the
button as well.
The authorization code may only be passed to specially trained personnel. You will
get the training and the authorization code by ASSKEA GmbH or an authorized
service partner of ASSKEA GmbH.
Passwords must be treated as confidential information to prevent misuse.
4.
After the authorization, the sensitivity setting of the alarm “System closed” is displayed.
The sensitivity can be adjusted with the
arrow buttons.
1
5
9
low sensitivity
default
high sensitivity
(delayed triggering of the alarm)
(rapid triggering of the alarm)
When the sensitivity is set to “1”, the device detects a full canister or a blocked
filter only.
For this setting, it is obligatory to reduce the interval time for checking the
dressing seal by the personnel to detect potential blockings in the aspirating
system promptly.
5.
Press the
button to confirm the entry.