3.1.5
Positioning of the ASSKEA pro
wound
®
S
The
ASSKEA pro
wound
®
S
can be placed next to the patient bed. Optionally, a variable holder for
attachment of the device to tube and rail systems as well as a bed holder are available.
To ensure optimum aspiration of the wound exudate, place the
ASSKEA pro
wound
®
S
below the wound to be treated. It should be noted that the suction tube does not form a
dip and is situated at least on patient level.
3.1.6
Connecting the ASSKEA disposable exudate canister system of the
ASSKEA pro
wound
®
S
ATTENTION!
Malfunction due to collapsing disposable liner “OneWay”.
A leak in the external canister “Bag” or at the lid of the disposable liner
“OneWay” may cause air to flow into the external canister “Bag”. This may lead
to the collapse of the disposable liner “OneWay”.
Inspect the disposable exudate canister system (1,000 ml) to ensure that the
lid of the disposable liner “OneWay” is firmly connected to the external
canister “Bag”.
Ensure that all connections are firmly attached and properly connected.
Ensure that the external canister “Bag” is undamaged and the T-piece is
firmly attached.
Follow the instruction for use supplied by the manufacturer!
The original ASSKEA disposable exudate canister system consists of the external canister “Bag”,
the holder for the external canister “Bag”, the disposable liner “OneWay”, the connecting tube for
the disposable liner “OneWay” and the sterile disposable suction tube with step connector.
Connection designation
A
Vacuum connection
B
Patient connection
Please also follow the instruction for use
supplied with the disposable exudate
canister system (1,000 ml)!
Fig. 6