3.1
Set-Up and Startup
3.1.1
Startup
It is important to follow the safety instructions in chapter 1.6 prior to initial startup.
Always have one backup disposable exudate canister for the
ASSKEA pro
wound
®
M
and one
backup disposable liner „OneWay“ for the
ASSKEA pro
wound
®
S
ready, since it is absolutely
necessary for safe operation!
Please read this entire instruction for use before operating the
ASSKEA pro
wound
®
M or
ASSKEA pro
wound
®
S
device for the first time.
Remove the device and the accessories from the packaging.
Always place the device on a sturdy and flat surface, take care of the correct position of the
device.
Inspect all tubing as well as the power supply unit for damage prior to each startup of the
ASSKEA pro
wound
®
M and ASSKEA pro
wound
®
S
. It is important to avoid kinking when
connecting the tubing. Ensure prior to switching on the unit that the disposable exudate
canister and tubings are properly connected.
Fully charge the battery prior to initial startup.
Perform a function test! (Please refer to chapter 5.1)
3.1.2
Connecting the ASSKEA pro
wound
®
M and ASSKEA pro
wound
®
S
Use the socket for power supply unit of the
ASSKEA pro
wound
®
M
(chapter 2.1.1, fig.1 (G)) or the
socket for power supply unit of the
ASSKEA pro
wound
®
S
(chapter 2.1.3, fig.2 (G)) to connect the
device to the mains power supply via the supplied power supply unit (type: FRIWO FW 7555M/12)
for charging or operation as required.
Use the supplied power supply unit only. First connect the power supply unit to the socket for
power supply unit of the
ASSKEA pro
wound
®
M or ASSKEA pro
wound
®
S
and then to the mains
power supply.
Fig. 3 Connecting the ASSKEA prowound
®
M and ASSKEA prowound
®
S to the patient and accessories