Operation manual
Ergometer CRG 200 v.5xx
IU-AT-77-C_I_ENG
ASPEL S.A.
19.05.2020
Strona 18 z 31
•
When using the Ergometer, place the power supply, power cable and data
cable in such a way that they are not exposed to mechanical damage.
•
After installing and starting the device, prepare an acceptance protocol
(signed by a person installing the device and the client) confirming that
the device has been appropriately installed and commissioned, according
to this manual and EN 60601-1 standard.
•
Apart from the acceptance protocol, the client should be provided with
Operation Manual for the Ergometer CRG 200 and computer set and
relevant declarations of conformity.
•
It is forbidden to conduct any modifications of location and composition
of the devices without prior written consent of the manufacturer.
•
After any modification in terms of device composition, compliance with
EN 60601-1: section 16 standard should be verified again.
•
Free space required for safe use of Ergometer should not be smaller than
0.6 m from the side with access to the device and must include the
landing area (free space can be shared if there is other training equipment
placed nearby).
•
In order to get off from Ergometer safely you should stop pedaling and
while holding the handlebar go down on either right or left side of the
device.
Fig.6. Free space required for safe use
>0,6 m
>0,6 m