SECTION 3 OPERATING THE A-2000
______________________________________________________________________
3-2
WARNINGS!
THE CONDUCTIVE PARTS OF ELECTRODES OR SENSOR AND
CONNECTORS, INCLUDING THE NEUTRAL ELECTRODE, SHOULD
NOT CONTACT OTHER CONDUCTIVE PARTS, INCLUDING EARTH.
TO REDUCE THE HAZARD OF BURNS IN THE HIGH-FREQUENCY
SURGICAL NEUTRAL ELECTRODE CONNECTION, THE SENSOR OR
ELECTRODES SHOULD NOT BE LOCATED BETWEEN THE
SURGICAL SITE AND THE ELECTRO-SURGICAL UNIT RETURN
ELECTRODE.
THE SENSOR MUST NOT BE LOCATED BETWEEN DEFIBRILLATOR
PADS WHEN A DEFIBRILLATOR IS USED ON A PATIENT
CONNECTED TO THE A-2000.
TO MINIMIZE THE RISK OF PATIENT STRANGULATION, THE
PATIENT INTERFACE CABLE (PIC) MUST BE CAREFULLY PLACED
AND SECURED.
3. Secure the DSC
Using the attachment clip, secure the DSC to a convenient location near the patient's
head.
4. Attach the BIS Sensor to the PIC
Figure 8 Connecting the PIC (BIS XP Platform and XP Compatible)
BIS XP Platform or XP Compatible:
To insert the sensor into the PIC, line up as
shown and insert the sensor tab into the PIC sensor connector until an audible
“click” is heard. The blank side of the sensor tab (i.e. the side without the computer
chip) should be facing up.
Standard System:
Line up the arrows and press firmly to ensure a good
connection.
For best results, use tape to secure the sensor in place. Adhesive should be used
only on the sensor, NOT on the PIC sensor connector.
PIC Sensor
Connector
Sensor tab
Release
Button