SECTION 3 OPERATING THE A-2000
______________________________________________________________________
3-18
BIS Log Interval (minutes)
– The interval at which the BIS values are reported in the
BIS Log display can be set to 1, 5, 15, or 60 minutes.
To set the BIS Log Interval:
1. Press
[MENU/EXIT]
to access the Setup Menu.
2. Highlight
Advanced Setup
using the up [
↑
] or down [
↓
] arrows.
3. Press
[SELECT]
.
4. Highlight
Display Parameter Setup
using the up [
↑
] or down [
↓
] arrows.
5. Press
[SELECT]
.
6. Highlight
BIS Log Interval (minutes)
using the up [
↑
] or down [
↓
] arrows.
7. Press
[SELECT]
until the desired interval appears in reverse video.
8. Press the up [
↑
] or down [
↓
] arrows to move to the next menu item, or
9. Press
[MENU/EXIT]
to exit.
Alternatively, when BIS Log display is active, press the up [
↑
] or down [
↓
] arrows to
increase or decrease the interval, respectively.
To permanently save this change, use the “Save Settings” option in the Advanced Setup
Menu. When ICU Mode is selected or when a BIS Extend sensor is attached, BIS Log
interval is automatically set to 60 minutes.
Return to Advanced Setup Menu -
Returns to the previous menu.
1. Press the up [
↑
] or down [
↓
] arrows to highlight
Return to Advanced Setup
Menu.
2. Press
[SELECT]
to re-display the Advanced Setup Menu.
3.4.5 The Diagnostic Menu
The Diagnostic Menu is covered in detail in Section 7.
3.4.6 Alarm
Menu
Alarm Menu
to Exit
Press
High Alarm Limit OFF
Low Alarm Limit OFF
Return to Advanced Setup Menu
Alarm Suspend ON OFF
Figure 18 The Alarm Menu