SECTION VII DISASSEMBLING AND REASSEMBLING THE A-2000™
- 7-26 -
7.10
THE CLAMP SHOE ASSEMBLY
7.10.1
Removing the Clamp Shoe Assembly
1. Open the monitor case as described in Section 7.2.1.
2. Remove the Interconnect PCB as described in Section 7.5.1.
3. Remove three screws securing shoe clamp assembly. (See Figure 7-20).
7.10.2
Installing the Clamp Shoe Assembly
1. Install screws securing clamp shoe assembly using Loctite 425.
2. Install the Interconnect PCB w/ Components as described in 7.5.2.
3. Close the monitor case as described in section 7.2.2.
Caution:
VERY IMPORTANT! Use only the Loctite number specified. Using
the incorrect Loctite will damage plastic components.