SECTION VI DIAGNOSTICS AND TROUBLESHOOTING
- 6-11 -
Table 6.4 Automatic Diagnostic Error Messages, Causes and Corrective Actions
(CONTINUED)
Failure Messages:
Causes:
Corrective Actions:
Battery
E33
Operating On Battery
Backup
The AC power has been
lost and the monitor is
running on the battery. The
battery keeps the monitor
operating for approximately
20 minutes.
Restore the AC power.
E34
Battery Power Low
There are only a few
minutes of battery life left.
Restore AC power to avoid
automatic shutdown.
Software
E35 - E39, E43 - E45, E48 -
E65
Software Error
A serious software error has
occurred. The monitor may
stop operating.
1. Turn the monitor off, then
on again.
2. Update the software.
Memory
E40
Setup Memory Write Error
Error in erasing or writing
Setup block of FLASH
memory.
1. Check Setup.
2. Replace main board.
E41
Real-Time Clock Write
Error
Error in writing to Real-Time
Clock RAM.
Replace main board.
E42
Can’t Save Setup
Monitor is unable to write to
Setup Memory when battery
power is low.
1. Restore AC power.
2. Save settings again.
E47
Data Memory Write Error
Error in erasing or writing
Data Memory portion of
FLASH memory.
Replace main board.
Summary of Contents for BISPECTRAL INDEX A-2000
Page 9: ...SECTION I INTRODUCTION 1 4 Figure 1 1 The A 2000 BIS Monitoring System...
Page 11: ...SECTION I INTRODUCTION 1 6...
Page 19: ...SECTION III PRINCIPLES OF OPERATION 3 2 Figure 3 1 The A 2000 System Block Diagram...
Page 20: ...SECTION III PRINCIPLES OF OPERATION 3 3 Figure 3 2 The A 2000 Signal Flow Diagram...
Page 35: ...SECTION V PREVENTIVE MAINTENANCE CARE AND CLEANING 5 6...
Page 53: ...SECTION VI DIAGNOSTICS AND TROUBLESHOOTING 6 18...
Page 61: ...SECTION VII DISASSEMBLING AND REASSEMBLING THE A 2000 7 8 Figure 7 4 The Four Mounting Screws...