SECTION V PREVENTIVE MAINTENANCE, CARE AND CLEANING
- 5-1 -
SECTION V
5.
PREVENTIVE MAINTENANCE, CARE AND CLEANING
INTRODUCTION
This section describes:
•
Care and cleaning procedures
•
Routine maintenance
5.1
CARE AND CLEANING
WARNING:
UNIVERSAL PRECAUTIONS SHALL BE OBSERVED TO PREVENT
CONTACT WITH BLOOD OR OTHER POTENTIALLY INFECTIOUS
MATERIALS. PLACE CONTAMINATED MATERIALS IN REGULATED
WASTE CONTAINER.
5.1.1
Cleaning the Monitor and Digital Signal Converter
Clean any spillage of blood or solutions on either the monitor or Digital Signal Converter
as soon as possible. Dried blood is very difficult to remove. Use lint-free absorbent
towels for spill cleanups. Dampen the towel with detergent and lukewarm water to aid in
cleaning.
5.1.2
Disinfecting the Monitor and Digital Signal Converter
Use lint free absorbent towels dampened with 10% bleach solution, or a commercial
disinfectant (e.g. Lysol Professional Disinfectant Foam Cleaner Spray or PDI Germicidal
Disposable Wipes).
WARNING:
WHENEVER AN EVENT SUCH AS SPILLAGE OF BLOOD OR
SOLUTIONS OCCURS, RE-TEST GROUND LEAKAGE CURRENT
BEFORE FURTHER USE.
DO NOT MIX DISINFECTING SOLUTIONS (e.g. BLEACH AND
AMMONIA) AS HAZARDOUS GASES MAY RESULT.
Caution:
Do not autoclave the Digital Signal Converter or Monitor.
Autoclaving will seriously damage both components.
Do not submerge.
After cleaning, dry all areas except the monitor display screen (see below) with a lint-free
absorbent paper towel.
5.1.3
Cleaning the Monitor Display
Clean the monitor display screen with a mild solution of detergent and warm water, or a
commercial display screen cleaner, available through personal computer dealers. To
avoid scratching the screen, never use abrasive cleaners.
Summary of Contents for BISPECTRAL INDEX A-2000
Page 9: ...SECTION I INTRODUCTION 1 4 Figure 1 1 The A 2000 BIS Monitoring System...
Page 11: ...SECTION I INTRODUCTION 1 6...
Page 19: ...SECTION III PRINCIPLES OF OPERATION 3 2 Figure 3 1 The A 2000 System Block Diagram...
Page 20: ...SECTION III PRINCIPLES OF OPERATION 3 3 Figure 3 2 The A 2000 Signal Flow Diagram...
Page 35: ...SECTION V PREVENTIVE MAINTENANCE CARE AND CLEANING 5 6...
Page 53: ...SECTION VI DIAGNOSTICS AND TROUBLESHOOTING 6 18...
Page 61: ...SECTION VII DISASSEMBLING AND REASSEMBLING THE A 2000 7 8 Figure 7 4 The Four Mounting Screws...
Page 91: ......