SECTION II SAFETY PRECAUTIONS
- 2-3 -
WHEN CONNECTING EXTERNAL EQUIPMENT (e.g. DATA CAPTURE COMPUTER),
THE SYSTEM LEAKAGE CURRENT MUST BE CHECKED AND MUST BE LESS
THAN THE IEC601-1-1 LIMIT.
THE USE OF ACCESSORY EQUIPMENT NOT COMPLYING WITH THE EQUIVALENT
SAFETY REQUIREMENTS OF THIS EQUIPMENT MAY LEAD TO A REDUCED LEVEL
OF SAFETY OF THE RESULTING SYSTEM. CONSIDERATION RELATING TO THE
CHOICE SHALL INCLUDE:
•
USE OF THE ACCESSORY IN THE PATIENT VICINITY;
•
EVIDENCE THAT THE SAFETY CERTIFICATION OF THE ACCESSORY HAS
BEEN PERFORMED IN ACCORDANCE TO THE APPROPRIATE IEC 601-1
AND/OR IEC 601-1-1 HARMONIZED NATIONAL STANDARD.
FAN WIRE MUST BE POSITIONED AS SHOWN IN FIG. 7-21. INCORRECT ROUTING
OF FAN WIRE MAY RESULT IN A HAZAROUS CONDITION.
2.1.3
Cautions:
U.S. Federal law restricts this device to sale by or on the order of a physician.
Refer to the section on disassembling and reassembling before attempting to
service the inside of the A-2000.
Carefully read the entire A-2000 Operating Manual before using the monitor in a
clinical setting.
Periodically check the battery by operating an A-2000 that has been disconnected
from the wall socket and which has been charging the battery for at least four
hours. After long periods of storage (e.g. > 1 month) it may be necessary to cycle
(discharge, then charge) the battery a few times to get full charge capacity. If the A-
2000 fails to operate reliably from the battery for twenty minutes, battery
replacement is required.
The A-2000 contains an internal Nickel-Metal-Hydride battery. The battery must be
removed by a qualified service technician and disposed of or recycled in
accordance with the national laws of the country. Contact Aspect Medical Systems,
Inc. or the local distributor for a replacement battery.
A burned out fuse usually indicates a serious problem with the electrical system of
the A-2000 unit. Call service before attempting to remove and replace a fuse.
If metal comes in contact with the battery terminals, permanent damage to the
battery and other connecting parts could result.
Do not autoclave the Digital Signal Converter or Monitor. Autoclaving will seriously
damage both components.
Do not submerge.
Do not block fan output. Keep at least three full inches of unobstructed space
between rear of instrument and wall and other instruments. Do not allow other
instruments to block airflow. Do not block ventilation inlet holes on the underside
of monitor.
Summary of Contents for BISPECTRAL INDEX A-2000
Page 9: ...SECTION I INTRODUCTION 1 4 Figure 1 1 The A 2000 BIS Monitoring System...
Page 11: ...SECTION I INTRODUCTION 1 6...
Page 19: ...SECTION III PRINCIPLES OF OPERATION 3 2 Figure 3 1 The A 2000 System Block Diagram...
Page 20: ...SECTION III PRINCIPLES OF OPERATION 3 3 Figure 3 2 The A 2000 Signal Flow Diagram...
Page 35: ...SECTION V PREVENTIVE MAINTENANCE CARE AND CLEANING 5 6...
Page 53: ...SECTION VI DIAGNOSTICS AND TROUBLESHOOTING 6 18...
Page 61: ...SECTION VII DISASSEMBLING AND REASSEMBLING THE A 2000 7 8 Figure 7 4 The Four Mounting Screws...
Page 91: ......