SECTION VII DISASSEMBLING AND REASSEMBLING THE A-2000™
- 7-12 -
6. Loosen L bracket screw (Fig. 7-10). Be sure to check that the alignment of bottom
screws is adequate, then install rubber foot, flat head steel (in center position), and
flat head nylon screw (right side). Tighten nylon screw first. Do not over-tighten nylon
screw. Then tighten L bracket screw.
Note:
It is important to follow the instructions regarding the sequence of
tightening the screws in step 6 above. Failure to do so may result in
excess electro-magnetic radiation from the DSC monitor interface cable.
7. Install two screws at A/C input module.
8. Tighten all screws and equipotential stud. Do not over-tighten.
9. Close case using the instructions in Section 7.2.2.