Gebrauchsanweisung LIANA

 

| Version 1.0 

 

Seite 11 von 50 

 
 

Environmental protection/waste disposal 

 

The manufacturer is aware of its responsibility towards the environment. The product is not to be disposed of 
with household waste! 

 

In accordance with WEEE, the manufacturer will take back all equipment placed on the market by him for the 
purpose of proper disposal. Please contact us in this regard if required and inform your customers in case of 
resale. 

  

 

When using the device together with other medical devices, take care not to trap or crush the patient's body or 
parts  of  the  patient's  body!  You  should  also  watch  out  for  instability/tipping  over  while  the  patient  is  being 
transferred! 

 

 

CAUTION

 

If any unusual noises, damage or malfunctions occur, operation of the product must be discontinued. 

 

 

NOTE

 

Repairs  to  parts  of  the  product may  only  be carried  out  by  trained  specialist  personnel  authorised  by  the 
manufacturer. Be sure to contact your customer service. 
Opening the device or accessories will void the warranty and any liability claims. 

 

 

WARNING

 

Any unauthorised repairs, conversions and modifications are prohibited for safety reasons and exclude the 
manufacturer's liability for any resulting damage. 
Any further liability of the manufacturer for damage resulting from the use of spare parts or accessories not 
approved by the manufacturer is excluded. 

 

 

CAUTION

 

The patient must not be left unattended at any time, in order to prevent injuries and falls or similar. 

 

 

CAUTION

 

This device may contain small parts that could be inhaled or swallowed, posing a choking hazard to young 
children. Keep children and pets away from the device. 
The wired remote control poses a strangulation risk. Take all the necessary precautions to prevent this. 

 

 

CAUTION

 

The stabilisation straps pose a strangulation risk. Take all the necessary precautions to prevent this. 

 

 

CAUTION

 

When attaching the T-element, always make sure that the latching elements engage on both sides and that 
the  gap  between  the  mattress  and  the  T-element  is  not  too  large  to  allow  the  patient  to  slip  through  -> 
Strangulation risk/risk of falling. 

 
 

Summary of Contents for motioncare LIANA

Page 1: ...TABLE OF CONTENTS Before starting up and using the medical product the user must become thoroughly familiar with this manual...

Page 2: ...onstructive assembly 27 10 Handswitch and column keypad 28 11 Operation 29 11 1 Switching the device on and off 29 11 2 Battery rechargeable battery 29 12 Parking position 30 12 1 Parking position of...

Page 3: ...Kalkofen 24 D 57439 Attendorn Fon 49 2722 63596 0 Fax 492722 63596 20 info asp d de www systems for life com ASP is certified by T V Rheinland according to DIN EN ISO 9001 2008 DIN EN ISO 13485 2016 D...

Page 4: ...to report serious incidents to the competent authority of the respective member state Other regulations may apply in other regions 2 2 Liability and warranty Through the information in this manual the...

Page 5: ...ety instructions before operating the product Please keep the operating instructions in a safe and easily accessible place so that you can refer to it at any time 3 2 Nameplate This illustration shows...

Page 6: ...sation aid Chassis with electrically adjustable lifting columns vertical horizontal adjustable armrests detachable adjustable footrests waist strap L 1000L1 L 1000R1 L 1000A1 L 1000LE1 L 1000H1 L 1000...

Page 7: ...t protected against the effects of major mechanical forces and only has limited protection against the penetration of splashing water into the housing motor control unit and battery The product may on...

Page 8: ...bol Medical device Fuse CE marking according to MDR Air pressure limits Lowest air pressure Highest air humidity Name and address of the manufacturer Air humidity limits Lowest air pressure Highest ai...

Page 9: ...rol box operate the safety switch on the unit or open the door to the battery Dry the wet spots on the unit manually In the event of excessive overheating disconnect the battery from the unit charging...

Page 10: ...ensure that the patient can be mobilised safely Do not stand between the LIANA and an obstacle during the transfer Make sure that the power supply is always switched on during use Never cover tape ove...

Page 11: ...service Opening the device or accessories will void the warranty and any liability claims WARNING Any unauthorised repairs conversions and modifications are prohibited for safety reasons and exclude...

Page 12: ...remove the cardboard proceed as follows Cut the tension strap with the cutter knife Remove the tension strap Lift the box upwards and set it aside 5 1 2 Remove the product from the pallet The product...

Page 13: ...responsibility for the impact on the safety reliability and performance of the device under these conditions If the product is newly connected the technical information must be observed NOTE Before i...

Page 14: ...n a approx 86 kg Weight of backrest system n a approx 34 kg Battery weight n a 2 9 kg Max load column stroke n a 250 kg Voltage of charging station n a 100 240 37 53 V VA Voltage output n a 24 maks 2...

Page 15: ...Gebrauchsanweisung LIANA Version 1 0 Seite 15 von 50 7 1 Product dimensions...

Page 16: ...roendoscopic videoendoscopic swallow examination 8 1 Sitting mobilisation sitting stabilisation Mobilisation at the edge of the bed Sitting on the edge of the bed 1 CHECK THE BED Before mobilising the...

Page 17: ...the edge of the bed behind the patient Once the correct position has been found the parking brakes on all 4 wheels can be operated 7 SET THE CORRECT HEIGHT The height of the backrest and arm structur...

Page 18: ...ing the lumbar support and backrest there is a risk of pinching Please exercise due care when lowering 10 HEADREST Adjust the headrest to suit the patient and fix it in place To change the position of...

Page 19: ...y adjusted both sideways and in length To adjust the armrests to the correct length loosen the lower star grip 7 and move the armrest forwards or backwards until the desired position is reached Repeat...

Page 20: ...Pull out the latching elements 10 Then insert the T element from the front Then lock the element accordingly When mounting the T element always make sure that the latching elements engage on the right...

Page 21: ...n of the sitting at the edge of the bed mobilisation process the patient is held from the front by a specialist The patient can be returned directly to the lying position ATTENTION Constant supervisio...

Page 22: ...ure that the safety T element is not mounted Then slide the rotated armrest back onto the arm strut Tighten the star grip 1 in the selected position Repeat this process on the other armrest To adjust...

Page 23: ...emove the leg supports from the parking position at the rear right and left of the device Inside the chassis there are guide rails 3 for attaching the leg supports Slide the leg support retaining clam...

Page 24: ...keypad Patient places arms on the armrests and grasps the handles The patient then assumes a ski jumper position by bending forward slightly 9 STANDING UP The patient is assisted by a medical special...

Page 25: ...completing the initial standing with LIANA the bed is lowered below buttock level This allows the patient to quickly sit back down on the bed after use 13 END THE STANDING PROCESS After the standing...

Page 26: ...uchsanweisung LIANA Version 1 0 Seite 26 von 50 14 BRAKES The four brakes can now be released 15 REMOVE LIANA Move the LIANA system away from the patient 16 RETURN THE PATIENT Lay the patient back int...

Page 27: ...justment of lumbar support 3 Armrest 13 Backrest adjustment 4 Handle 14 Headrest adjustment 5 T piece locking 15 Device column 6 Safety T piece T element 16 Personnel handle 7 Braked castor 17 Electri...

Page 28: ...ators provide information about the charge level and indicate when maintenance is due or if the safe working load has been exceeded 3 bars Battery is full no charging required remaining capacity 75 2...

Page 29: ...een pressed Note Pressing the safety switch and opening the door to the battery interrupts all electrical functions 11 2 Battery rechargeable battery If the battery charge is low after checking charge...

Page 30: ...ation These can be parked safely at the rear of the chassis on the right and left To do this push the fixing clamps 1 of the leg rests onto the corresponding guide rail 2 Secure the clamp via the star...

Page 31: ...ting stabilisation in combination with feeding In all other cases you can conveniently park the table by the device To do this place the table in the parking elements provided 4 and turn the lever 5 t...

Page 32: ...f the controller box 13 1 Cleaning instructions Clean the device and upholstery with a soft lint free cloth moistened with soapy water or a commercially available plastic cleaner To remove grease stai...

Page 33: ...refully disinfected between each use For manual surface disinfection disinfectants for medical devices or a disinfectant aerosol spray commonly used in the industry can be used Recommended surface dis...

Page 34: ...rooms and systems of a special nature DIN VDE 0100 Group 700 This DGUV test is only mandatory for Germany Other countries may have different regulations that must be observed WARNING Do not perform an...

Page 35: ...17 DISPOSAL OF THE PRODUCT The expected service life of the LIANA patient sitting standing stabilisation aid is approx 10 years At the end of the product s service life contact your motioncare contra...

Page 36: ...attery Operation via the control box control panel Castors make loud noises Clean the rollers and replace if necessary The device makes unusual noises Please inform Customer Service Damage to the unit...

Page 37: ...y persons not trained and authorised by the manufacturer unoriginal spare parts have been used expires in the event of negligence in any form To enforce your warranty claim send the product with the w...

Page 38: ...edical incident accident immediately inform ASP GmbH Zum Kalkofen 24 57439 Attendorn Deutschland Telefon 49 0 2722 63596 0 E Mail info asp d de and the Bundesinstitut for Arzneimittel und Medizinprodu...

Page 39: ...Gebrauchsanweisung LIANA Version 1 0 Seite 39 von 50 21 ELECTRICAL DIAGRAM...

Page 40: ...Gebrauchsanweisung LIANA Version 1 0 Seite 40 von 50...

Page 41: ...or a business unit not an independent company Distribution by ASP GmbH Zum Kalkofen 24 D 57439 Attendorn Germany Phone 49 2722 63596 0 Mon Thu 8 00 a m 4 00 p m Fri 8 00 a m 3 00 p m E Mail info asp d...

Page 42: ...Gebrauchsanweisung LIANA Version 1 0 Seite 42 von 50 23 CE DECLARATION OF CONFORMITY PSS LIANA...

Page 43: ...cushion L 1000G1 Body strap L 1000LE2 Lumbar strap L 1000F1 Footrest L 1000FS1 Standard rail pair L 1000KL1 Fixing clamps L 1000T1 Table L 1000TE1 T element L 308 02 Handle cover sterile disposable ma...

Page 44: ...specified below The customer or the user of the product should ensure that it is used in such an environment Transmission measurement Agreement Electromagnetic Environment Guidelines High frequency R...

Page 45: ...y voltage should be that of a typical business or Match hospital environment Overvoltage IEC 61000 4 5 1 kV Line against line 2 kV Line against grounding 1 kV Line against line 2 kV Line against groun...

Page 46: ...romagnetic location measurement a should be lower than the CONFORMITY LEVEL in all frequency ranges be b In the environment of devices that have the following symbol disturbances are possible NOTE 1 A...

Page 47: ...depending on the transmission frequency in m 150 kHz to 800 MHz d 0 35 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 04 0 12 0 23 0 1 0 11 0 38 0 73 1 0 35 1 2 2 3 10 1 1 3 8 7 3 100...

Page 48: ...lisator LIANA Manufacturer ASP GmbH Zum Kalkofen 24 D 57439 Attendorn Serial number _______________________________________ Date of purchase _______________________________________ Location __________...

Page 49: ...Gebrauchsanweisung LIANA Version 1 0 Seite 49 von 50 27 REPEAT INSPECTION REPAIR DGUV 3 SAFETY INSPECTION ETC Type of examination Date Result Measure Signature...

Page 50: ...Gebrauchsanweisung LIANA Version 1 0 Seite 50 von 50...