Gebrauchsanweisung LIANA
| Version 1.0
Seite 10 von 50
Functional safety
▪
When transporting the device to the patient’s bed, make sure that it does not collide with other equipment.
▪
The product may only be used or operated by instructed personnel.
▪
Only use under supervision. The patient must never be left unattended.
▪
The patient must not be the user/operator!
▪
When using stabilisation straps, the safety T-piece (T-element) (see page 12, item 6) must always be
attached -> Risk of strangulation/falling.
▪
The device should be kept under supervision at all times to prevent the stabilising mechanisms from
accidentally crushing the patient.
▪
Manual emergency function: If the column is no longer able to be moved in the vertical direction in the event
of a power failure or problems with the control system, it is possible for 2 people to lift it using the arms of the
arm supports. The weight of the front frame is approx. 34 kg.
▪
Before each use of the product, the user/operator must be convinced that the product is functionally safe and
that it and its accessories are in their proper condition (visual inspection, functional test, etc.). Before use,
check that the electrical functions are working properly.
▪
Furthermore, check the stabilisation straps used for visible damage. In the event of damage or malfunctions,
do not continue to use the product or accessories. There is a risk of injury!
▪
This medical device should only be used with original accessories according to Chapter 24.
▪
Plan each transfer carefully; only then can you adequately protect yourself and the patient.
▪
Avoid slippery surfaces and doorsteps and do not move the product on sloping or uneven floors.
▪
Fix the castors of the hospital bed, stretcher, etc. in order to ensure that the patient can be mobilised safely.
▪
Do not stand between the LIANA and an obstacle during the transfer.
▪
Make sure that the power supply is always switched on during use.
▪
Never cover, tape over or change slots and openings on the device.
▪
The housing of the medical device must not be opened! The product does not have any parts to be serviced
by the user. Never insert a foreign object into the product.
▪
Do not perform any repairs or modifications on the product! Otherwise, the correct functioning of the product
and safety may be jeopardised. This will also invalidate your warranty claims! Repairs may only be carried out
by trained specialist personnel authorised by the manufacturer. Be sure to contact motioncare's customer
service or that of its authorised distributors.
▪
The product must be disinfected after each use.
▪
Cleaning and disinfection are only permitted if the patient is not present!
▪
Caregivers must protect the skin and eyes from concentrated disinfectant and detergent. Gloves should be
worn to protect the skin. A mouth covering should be worn as protection against aerosols. Safety goggles
should be worn to protect the eyes.
Maintenance
▪
To ensure hazard-free operation, the product must undergo an annual inspection, safety inspection and
maintenance by persons authorised by the manufacturer who meet the requirements of the Medical Devices
Operator Ordinance! In case of doubt, contact your supplier or manufacturer. Failure to do so may result in
injury or an unsafe product.
▪
Installation, maintenance, servicing and testing activities are only permitted if the patient is not present!
▪
The product contains removable parts. Follow Chapter 9 on how to correctly identify parts to avoid the risk of
confusion and malfunction.
▪
The product contains non-removable parts. If necessary, these may only be replaced by personnel authorised
by the manufacturer. Please observe the assembly instructions.
Summary of Contents for motioncare LIANA
Page 15: ...Gebrauchsanweisung LIANA Version 1 0 Seite 15 von 50 7 1 Product dimensions...
Page 39: ...Gebrauchsanweisung LIANA Version 1 0 Seite 39 von 50 21 ELECTRICAL DIAGRAM...
Page 40: ...Gebrauchsanweisung LIANA Version 1 0 Seite 40 von 50...
Page 42: ...Gebrauchsanweisung LIANA Version 1 0 Seite 42 von 50 23 CE DECLARATION OF CONFORMITY PSS LIANA...
Page 50: ...Gebrauchsanweisung LIANA Version 1 0 Seite 50 von 50...