47
to set TimeF to 10ms, and press
key to confirm.
9
Use
,
keys or knob to set Mode value to
pulse
, and
press
key to confirm.
►
Mode:
◄
Puls
►
10
Press
key to exit the transient operation menu.
0.000V 0.000A
1.000A
t
CCH OFF
11
Press
key to turn on the load.
0.000V 0.000A
1.000A
t
CCH ON
12
Send the command “INIT”or”INIT:CONT” to the load.
13
One trigger occurs when pressing
+
key or the signal
at the trigger input terminal (TRIG IN) is low level.
0.000V 0.000A
5.000A
t
CCH ON
For the above operating procedures, the corresponding SCPI commands should be:
MODE
CCH
;
Sets the mode
TRAN
ON
;
Enables the transient operation
CURR:LOW
1
;
Sets a value to low level
CURR:HIGH
5
;
Sets a value to high level
TRAN:HTIME 200ms
;
Sets a value to high level time
TRAN:RTIME 10us
;
Sets a value to time for rising edge
TRAN:FTIME 10us
;
Sets a value to time for falling edge
TRAN:MODE PULS
;
Selects pulsed transient operation
INP
ON
;
Turns on the load
INIT
;
Initialize the trigger
TRIG
;
One trigger occurs
4.7.3 Toggled Transient Operation
Example 1: Assume that the load is in external triggering mode, set the load resistance value to
periodically switch between 200
Ω
and 500
Ω
; set the time for rising edge (TimeR) to 10ms; set time
for falling edge (TimeF) to 10ms; the load is in toggled transient operation, and a trigger occurs at this
time.
Procedures Operation
Descriptions
Display
1
Press
key to enter into the mode selection and parameter
setting menu; use
,
keys to select CRH mode; press
key to confirm and exit the mode selection and parameter
setting menu.
0.000V 0.000A
20.000
Ω
CRH OFF
2
Press
key to enter into transient operation. The display
shows “t” to indicate in CRH status.
0.000V 0.000A
2000.0
Ω
t
CRH OFF