18
Then press
key to turn on the input;
Or via remote command to set:
SCPI Command
Description
TRAN ON
Enables transient operation
CURR:LOW 5
Sets transient current low level to 5A
CURR:HLGH 10
Sets transient current high level to 10A
TRAN:LTIM 500us
Sets transient low level time to 500us
TRAN:HTIM 500us
Sets transient high level time to 500us
TRAN:RTIM 200us
Sets the time for transient rising edge to 200us
TRAN:FTIM 200us
Sets the time for transient falling edge to 200us
TRAN:MODE CONT Selects continuous operation
INPUT ON
Turns on the input
Figure 2-6 shows the current waveform of the load: the load’s input current reaches the transient
high level (10A) after 200us duration of rising edge, and remains at 10A for 500us. Then after 200us
duration of falling edge, the input current reaches the transient low level (5A), and remains at 5A for
500us. Repeat it in cycles.
Fig. 2-6 Continuous Transient Operation
2.4.2
Pulsed Transient Operation
The trigger function is required for pulsed transient operation. When there is no trigger occurs,
the load remains at the transient low level. After a trigger has been received, a pulse with three stages,
namely rising edge, transient high level, and falling edge, will appear, then the load returns to the
transient low level again. The
associated
parameters such as transient low level (LevelL), transient
high level (LevelH), high level time (TimeH), time for rising edge (TimeR), time for falling edge
(TimeF), and pulsed transient mode can be set through transient operation menu or via remote
command (CURRent:LOW <NRf+>, CURRent:HIGH <NRf+>, VOLTage:LOW <NRf+>,
VOLTage:HIGH <NRf+>, RESistance:LOW <NRf+>, RESistance:HIGH <NRf+>,
TRANsient:HTIMe <NRf+>, TRANsient:RTIMe <NRf+>, TRANsient:FTIMe <NRf+>,
TRANsient:MODE PULSe, TRANsient ON | OFF).
In order to get a pulse, an explicit trigger is required. The trigger can be an external trigger signal
received via the TRIG input on
the front panel, pressing the
key +
key, or ,the remote