2. Installation
2.1. Sensor installation
Sensor dimensions can be found on chapter 8.
Proper sensor installation is extremely important in order for your flowmeter to work correctly. Below, you
will find the minimum sensor installation requirements that need to be respected at all time.
Sensor
installation
requirements:
All MAGS1 sensors are supplied with 2 built in earthing electrodes that are sufficient for all applications with
metal pipes and tanks. However on applications where all pipes and tanks are manufactured from plastic, it
is recommended that earthing rings are also installed to ensure the maximum resistance of the sensor to
earth is <1 ohm.
Conductive (metal) pipe
Non-conductive (plastic, lined)
pipe