MAGS1 User Guide
17
V1.5 22-1-2016
4.14. User settings
To enter this section, it is necessary to enter the User Password “1111”.
Modbus
register
Modbus
address
No.
of
bytes
Data
type
No. of
decima
l
Min
Valu
e
Max Value
Default
Description
Read/
Write
2000
1999
4
Bool
0
0
1
1
Measurement,
0=Run,
1=Stop
R/W
2002
2001
4
Bool
0
0
1
1
Air Detector,
0=ON,
1=OFF
R/W
2004
2003
4
Longint
3
0
999
188
Air Constant
R/W
2006
2005
4
Longint
0
1
30
15
Samples per Avg.
R/W
2008
2007
4
Longint
0
0
5
3
Low Flow Cutoff,
0=OFF,
1=0.5%,
2=1%,
3=2%,
4=5%,
5=10%
R/W
2010
2009
4
Bool
0
0
1
0
Invert Flow,
0=No-invert,
1=Invert
R/W
2012
1011
4
Longint
3
0
36 000 000 3 600
Flow Qn
R/W
2014
2013
4
Bool
0
0
1
0
Electrode Cleaning R/W
2016
2015
4
Bool
0
0
1
0
Delete Aux +
Volume
R/W
2018
2017
4
Longint
0
1
247
1
Modbus Slave
Address
R/W
2020
2019
4
Longint
0
0
115200
9600
Modbus BaudRate,
9600,
14400,
19200,
38400,
57600,
115200
R/W
2022
2021
4
Longint
0
0
3
3
Modbus Parity,
0=Even, 1 stopbit,
1=Odd, 1 stopbit,
2=None, 2 stopbits,
3=None, 1 stopbit
R/W
2024
2023
4
Longint
0
0
60
15
Start Delay
R/W
Measurement
–
1 = Stop
– the unit shows actual flow, but the totalizers are stopped.
0 = Running
– totalizers are active.
Air Detector
–
Empty pipe detection - if the Air detector is active and the pipe is empty, the unit
automatically turns down the excitation and detect error.
Air Constant
– constant value to determine the Empty pipe detection limit.
Samples per Avg.
– the number of samples that the flowmeter will use for calculation of its displayed
average flow value.
Low Flow Cutoff
–the minimum flow the flowmeter will react on. Setting is done in percentage of Qn.
Invert Flow
– possibility to change direction of the flow displayed.