Electromagnetic Flowmeter M-910
18
Operation manual
All flowmeters and computer are connected parallel using twisted pair cable. At each end of the communications
line should be 470
Ω
terminations.
Interconnection of three flowmeters and computer using RS485 bus
Flowmeters are marked with numbers. These numbers are equal to the flowmeters RS485 address.
Program Flow910 is designed for flowmeter control using RS485 or RS232 serial bus. It can be ordered as M900
option.
Note: Communication through the serial port RS485 is a half duplex type. The flowmeter is a listener and sends
data only after a query from a computer. Each flowmeter has it’s own RS485 address. The range of
addresses is 0 to 255. Factory setting of RS485 address is 0. Communication speed is selectable between
1200 and 19200 Bd. For cables over 100m or if there is a noisy power supply voltage (especially peaks
generated, usually by motors, etc.), select communication speed below 9600 Bd.
Note: Serial port RS485 is galvanically isolated from other electronic circuits.
910 _ 1
910 _ 2
910 _ 3
PC
470
Ω
470
Ω