43
Electromagnetic Compatibility
00
1.
14
15
5.3
3
r
ev
. 8
Electromagnetic Immunity
Guidance and Manufacturer’s Declaration
Electromagnetic Immunity - For all Equipment and Systems
The
Maxi Sky 600
is intended for use in electromagnetic environment specified below. The customer or the user of the
Maxi Sky 600
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±8 kV contact
1
±15 kV air
1
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
±2 kV for power supply lines
±1 kV for input/output lines
Mains power quality should be that of
a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential mode
±2 kV common mode
Mains power quality should be that of
a typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5%
U
T
(>95% dip in
U
T) for 0.5 cycle
40%
U
T
(60% dip in
U
T) for 5 cycles
70%
U
T
(30% dip in
U
T) for 25 cycles
<5%
U
T
(>95% dip in
U
T) for 5 sec.
<5%
U
T
(>95% dip in
U
T) for 0.5 cycle
40%
U
T
(60% dip in
U
T) for 5 cycles
70%
U
T
(30% dip in
U
T) for 25 cycles
<5%
U
T
(>95% dip in
U
T) for 5 sec.
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the
Maxi Sky 600
requires continued
operation during power mains
interruptions, it is recommended that
the
Maxi Sky 600
be powered from an
uninterruptible power supply or a
battery.
NOTE: UT is the a.c. mains voltage
prior to application of the test level.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercials or hospital environment.
NOTE:
U
T is the AC mains voltage prior to application of the test level.
Notes:
1
The EUT was tested at 8 kV contact discharge and 15 kV air discharge as per client request.
Summary of Contents for Maxi Sky 600