28
Care and Maintenance
00
1.
14
15
5.3
3
r
ev
. 8
Inspections by an Authorized Service Technician
FREQUENCY
Inspections for spreader
bar and slings
Initially
Before
every
use
Every two
months or
500 cycles
Every four
months or
1000 cycles
Every year
or
2500 cycles
Every two
years or
5000 cycles
Inspect all sling parts
(attachments, fabric, stitch
areas and strap) for signs
of wear, discoloration,
deterioration or loose
threads.
X
Clean sling as indicated
on the tag.
When necessary
Inspect the spreader bar
on the strap of the lift for
damage or cracks. Make
sure all attachments are
properly secured (e.g. split
ring).
X
FREQUENCY
Inspection for lift
Initially
Before
every
use
Every two
months or
500 cycles
Every four
months or
1000 cycles
Every year
or
2500 cycles
Every two
years or
5000 cycles
Replace strap.
X
Inspect frame parts
interlock and hardware for
malfunction and make
sure there are no parts
missing.
X
Inspect gears for wear.
X
Inspect connecting joints
for proper attachment
(trolley and spreader bar).
X
Verify the emergency
brake.
X
Verify emergency lowering
mechanism.
X
Verify alternative UP and
DOWN buttons on
cassette.
X
Load test (safe working
load -SWL) for ceiling lift
recommended.
X