6
The equipment must be used in accordance
with these safety instructions.
Anyone using the equipment must also have
read and understood the instructions in this
manual. If there is anything you are not sure
about, contact your local ArjoHuntleigh
representative.
Safety Instructions and Warnings
•
The Semi-Permanent Rack system
installation must be performed by two
persons.
•
All track ends must have a KWIKtrak
autolock stopper in place and secured.
•
The Semi-Permanent Rack system is
designed to be used with all KWIKtrak
lifts having a service weight load rated
to a maximum of 600 lb.
•
Make sure all components are present
(see table on page 7) .
•
Follow the assembly procedure
sequence.
•
Use each and every one of the parts –
DO NOT leave any out.
•
This equipment must be used by a
caregiver with proper training to work
with the patient to be transferred.
•
When performing a transfer, do not
allow the patient to swing on the rack.
WARNING: The maximum track length for
this system is determined by the lift in use,
as well as the track size. See Fig. 1. for
details.
Fig. 1
Maximum Mobile Track Span VS
Lift Capacity & Track Style
Track
Lift
Capacity
90 mm
140 mm
180 mm
200 Kg
(440 lb)
3 m
(9 ft 10 in)
4.5 m
(14 ft 9 in)
6 m
(19 ft 8 in)
270 Kg
(600 lb)
3.5 m
(11 ft 6 in)
4.5 m
(14 ft 9 in)
WARNING: Firmly tighten all fasteners.
Periodically inspect and retighten fasteners
to a torque of 20 N•m (15 lbf•ft) as required.
Safety Instructions