10. TROUBLE SHOOTING
Symptom
Possible Cause
Action
All actuators fail to work
Power disconnected. or
batteries discharged
Ensure mains supply lead is
connected
Control box and handset
plugs disconnected
Check the plugs are fully
inserted in the control box
sockets.
Blown fuse
Contact your local Arjo
Service Department.
Damaged handset cable
Replace handset
One actuator fails to work
Actuator plug disconnected
Check that actuator plug is
fully inserted in control box.
Main lift actuators fail to work,
and an audible beeping sound
can be heard when handset is
activated
Control box requires
calibrating possibly due to bed
being operated when batteries
were fully discharged
Ensure mains power lead is
connected and carry out
calibration procedure.
Brake lock not effective
Wear or damage
Adjust or replace castor