background image

Page | 18 

 

Commercial in confidence. Instructions For Use Revision SOR-REV5-090921

 

 

9.  MAINTENANCE

 

 

The following preventative maintenance procedures should be carried out every 12 months unless 
otherwise specified. 
 

Note: 

Other than general maintenance activities all other service, or repair work must be performed 

by qualified and trained persons who are approved by Arjo. 

 

WARNING

 

The bed must be disconnected from the mains power supply before starting any 
maintenance activity. 

 

GENERAL 

Examine all the warning signs on the bed, examine the bed for obvious signs of 
damage. All aspects of the equipment should operate as intended. Check that all nuts, 
bolts and other fasteners are tight and are not missing. 

Examine flexible cables and conduits for cuts, cracks, abrasions, or other deterioration. 

Check that the power supply plug is not damaged. If either the power cable, or plug is 
damaged then both the cable and plug must be replaced as a complete assembly by 
Arjo, or an approved service agent. 

CASTOR BRAKING/LOCKING SYSTEM 

Apply the brakes and push the bed forwards. If any of the four braking castors rotate 
the brake is not fully effective. 

SIDE RAILS (if fitted)

 

  The locking pin mechanism and striker plate arrangement should be checked 

every three months. 

  Inspect the side rails for general wear and tear on fittings and pivot joints every 

six months. 

  Check for rust, structural integrity of welds and paint finish every twelve 

months. 

Raise the side rail to the upright position until the locking mechanism clicks into place. 
Check that the there is no yellow shaft visible. In the locked position, the yellow on the 
locking pin shaft should not be visible from above. Refer to the operating procedures on 
page 12 of this manual for further information. 

If the yellow shaft is visible when the side rail is in the raised and locked position, 
contact Arjo for service advice. 

Check that the rails pivot axles are securely fitted to the bed. The pivot axles are locked 
into place by the manufacturer using a 10mm lock nut. It is rare that re-tightening of the 
axles would be required, but in the case that it is, ensure that the nut is not over 
tightened during the tightening procedure. 

 

Summary of Contents for Sorrento Floor Bed

Page 1: ...Instructions for Use Single King Single Bed Options Sorrento Floor Bed SOR REV5 090921...

Page 2: ...opulmonary Resuscitation CPR 8 4 9 Lock Out Function 8 4 10 Battery Charge Indicator 8 4 11 Mains Power Indicator 9 4 12 Under Bed Light 9 5 GENERAL BED OPERATION 10 14 5 1 Castor System 10 5 2 Bed Ex...

Page 3: ...iate for long term adult patient care under supervised medical monitoring and provides support with various positioning options Key Features Total safe working load SWL of the bed 250Kg Four 75mm brak...

Page 4: ...equipment can be hazardous if used incorrectly Do not use electrically operated beds in the presence of flammable gases Use the bed to support one patient only The Sorrento bed is intended for one pe...

Page 5: ...eft in the flat position whilst unattended The clinically qualified person responsible should consider the age size and condition of the patient before using the bed and its accessories Always ensure...

Page 6: ...Use Revision SOR REV5 090921 3 ELECTRONIC FUNCTIONS 3 1 Handset Foot Board Control Panel Operation The Sorrento s bed adjustments are powered by four electric actuators which are controlled by the but...

Page 7: ...ctuator force in the event of being obstructed in lowering 4 2 Mattress Platform Height Adjustment WARNING It is recommended that the bed be positioned at its lowest height when unattended by caregive...

Page 8: ...t or head down tilt position Tilt positioning achievable is between 0 and 14 for both head and foot down tilting In use the mattress platform may initially rise before tilting The bed will automatical...

Page 9: ...on the control panel by LED lights in between the function buttons To lock and unlock each function press the lock button and at the same time press the function button that is required to be locked...

Page 10: ...l devices fitted to the bed The under bed lighting feature provides a soft LED illumination of the floor beneath the bed carriage This safety feature provides floor illumination for patients at night...

Page 11: ...ase the braking castors use the pedals at each side and end of the bed Simply push down on the red lever using the toe of the shoe to brake and the green lever to release the brake CAUTION Disconnect...

Page 12: ...screw the handwheels both sides of the bed anti clockwise until approximately 20mm of thread is showing Pull the foot panel towards you using two hands evenly spaced on the panel Pull the panel to its...

Page 13: ...et these requirements when fitted correctly 5 4 Fitting the Side Rails NOTE The rails are a left and right hand set and when fitted correctly fold down towards the foot end of the bed 1 Before fitting...

Page 14: ...rs visitors and or other equipment Side Rail Operation To raise the side Lift the top rail until the magenta locating pin knob automatically clicks into its locked position at maximum height To lower...

Page 15: ...ellow shaft visible If there is yellow shaft visible the safety side is NOT securely locked in the upright position and needs to be adjusted The yellow coating should NOT be visible from above when th...

Page 16: ...Safe Working Load Safety Factor 2 x S W L 250 Kg Maximum Patient Weight 215 Kg Electronic Parameters Power Input 100 240V AC 50 60Hz Max current 3 9 A Rate Load interruption 10 Max 2 minutes 18 minute...

Page 17: ...ely Safe Working Load Safety Factor 2 x S W L 250 Kg Maximum Patient Weight 215 Kg Electronic Parameters Power Input 100 240V AC 50 60Hz Max current 3 9 A Rate Load interruption 10 Max 2 minutes 18 mi...

Page 18: ...hot water and mild detergent or the facilities recommended cleaning solution It is recommended that the cloth be a white material to prevent bleeding of die colours to the bed Take extra care in areas...

Page 19: ...ASTOR BRAKING LOCKING SYSTEM 1 Apply the brakes and push the bed forwards If any of the four braking castors rotate the brake is not fully effective SIDE RAILS if fitted 1 The locking pin mechanism an...

Page 20: ...dset is activated Control box requires calibrating possibly due to bed being operated when batteries were fully discharged Ensure mains power lead is connected and carry out calibration procedure Brak...

Page 21: ...he product must be recycled separately Batteries are to be disposed in accordance with national or local regulations Components that are primarily made up of different kinds of metal containing more t...

Page 22: ...Page 21 Commercial in confidence Instructions For Use Revision SOR REV5 090921 14 SORRENTO BED LABELS WARNINGS...

Page 23: ...abel Areas labelled are of a low risk nature Recommended Mattress Size Label Manufacturer European Representative Date of Manufacture Product Code Reference Serial Number Bed Safe Working Load Maximum...

Page 24: ...ions which provides the full details of warranty terms and statutory rights of the customer can be obtained upon request 17 REGULATORY INFORMATION Caretek China Medical PLC Xiguakeng Guanqiao Village...

Reviews: