background image

001.10610.EN rev

. 13

Assembling the EASYTRACK

9

b)  Press in metal pins 

on both sides of the 

post. Metal pins 

cannot be pressed 

in when there is 

downward pressure 

on them. Pressure 

must be released to 

press in pin.

Fig. 15 

c)  Pull up the top portion of the post 

towards the ceiling.

d)  Repeat until the top of the post is less 

than 76 mm / 3” from the ceiling.

9)  Lift the blue locking 

handle out of the lower 

portion of the post (see 

Fig. 2 - item № 4). Using 

the blue piece as a 

handle, twist the post as 

shown in Fig. 16:

a)  Top portion – turn 

counter-clockwise.

b)  Bottom portion – 

hold firmly.

Fig. 16 

10) Before the EASYTRACK 

post tightens against the 

ceiling; check the clip-on 

level to make sure the 

bubble is in the centre of 

the circle. Tilt the post 

until the bubble reaches 

the centre of the circle, 

see Fig. 17 and Fig. 18.

Fig. 17 

NOTE: 

Make sure clip-on 

level is fully attached to 

the post. Failure to attach 

level properly could give 

an incorrect reading. Level 

should “Click” into place.

Fig. 18 

11) Once the bubble is in the centre of the clip-

on level – continue to tighten 

only until the 

red mark at the top of the post is no 

longer visible

.

Fig. 19

Incorrect

Correct

Red mark

visible

Red mark 

not visible

WARNING

: Red mark 

must not be visible for 

the post to be safely 

secured (see the warning 

label, apposed on each 

post, as shown in 

Fig. 20.

NOTE: 

Do not over-

tighten the post.  

Tighten only until red 

mark disappears.  

Over-tightening post  

may cause damage to 

the ceiling.

Fig. 20 

12) Remove the level from 

the post by sliding it up 

off the bottom portion of 

the post. Slide the blue 

locking device down to fit 

into the bottom part of 

the post to lock the post 

in place.

Fig. 21 

NOTE: 

Once the blue locking handle is 

pressed into the lower portion of the post, 

the post can be turned as a whole, except 

for the top plate and foot plate. This will not 

lower the post. The post must turn to enable 

you to attach the track.

13) Repeat procedure from steps (1) for the 

second and third posts.

NOTE: 

Make sure that all posts are far 

enough from each other to allow the track to 

latch on (minimum distance between each 

post: 1.9 m (75”).

14) Get the Fixed track with 

the swivel trolley (Fig. 2 

–  item  №  11  &  12)  and 

place the end of the track 

marked  “Step  1”  onto 

one standard post, not 

the post designated to 

be the ‘pivot’ post. Place 

the track onto the pins 

and rotate the post to 

desired direction.

Fig. 22 

NOTE: 

Tilt the track upward (45°) to latch 

the track onto the post.

Summary of Contents for EASYTRACK 9330010

Page 1: ...001 10610 EN rev 13 10 2019 INSTRUCTIONS FOR USE EASYTRACK SYSTEM 3 POST ASSEMBLY Model no 9330010...

Page 2: ...ing to the Arjo group of companies Arjo 2019 As our policy is one of continuous improvement we reserve the right to modify designs without prior notice The content of this publication may not be copie...

Page 3: ...ice 3 Homecare Environment Considerations 3 Symbols Used 4 General Symbols 4 Product Description 5 EASYTRACK System Requirements 6 Ceiling Requirements 6 Floor Requirements 6 Other Requirements 6 Room...

Page 4: ...Page intentionally left blank...

Page 5: ...Arjo vendor If a serious incident occurs in relation to this medical device affecting the user or the patient then the user or patient should report the serious incident to the incident to the medica...

Page 6: ...part and lead to a patient fall The life span is the safe operating time period before a complete overhaul is required Following are factors that have an impact on the EASYTRACK life span aging of th...

Page 7: ...r Do not drop product may cause breakage Do not completely submerge in water Do not attempt to over extend the track past the locking point Do not install product if Neoprene on top or bottom plate ap...

Page 8: ...5223 3 15 Points out the manufacturer s catalogue number Indicates a potential risk see Instructions for Use for corrective action SN Points out the manufacturer s serial number SWL SWL is the maximum...

Page 9: ...5 Post top for pivot post 700 10560 1 6 Top plate 700 10400 3 7 Pivoting Track extendable 1 9 3 m with the standard trolley 700 10850 1 8 Standard trolley 700 10650 1 9 Standard post 700 10500 2 10 C...

Page 10: ...t fall Floor Requirements Flooring must comply to local building code standards WARNING Termites or water damages may affect the structure of the floor Floor must be clear of any dust grease water or...

Page 11: ...the three posts considering the following see Fig 6 Distance from the wall adjacent to the two standard posts to allow clearance for the pivot track overhang Minimum distance of 1 9m 75 between the t...

Page 12: ...ps into place Fig 10 5 Take one of the posts out of the box Turn the post upright Place the clip on level see Fig 2 item 16 onto the post at approximately waist height Fig 11 6 Place the post onto the...

Page 13: ...longer visible Fig 19 Incorrect Correct Red mark visible Red mark not visible WARNING Red mark must not be visible for the post to be safely secured see the warning label apposed on each post as show...

Page 14: ...f the pivot post Tilt the track upward 45 to latch it onto the post Fig 28 21 Extend the track and thread the wheels of the trolley through the upper portion of the track Fig 29 22 Continue to extend...

Page 15: ...lation against Fig 36 Make sure the red mark on top of each posts is not visible Using the clip on level make sure that all posts are straight Make sure that the both tracks are locked into place and...

Page 16: ...ged from the trolley tilt it to a 45 angle and lift the end at the pivoting post to release it from the hooks Fig 42 9 Put the end stopper back on the track and secure it by tightening the Allen key 1...

Page 17: ...essed in when there is downward pressure on them Pressure must be released to press in pin d Push top portion of post downward e Repeat until the metal pins are no longer visible and the top of the po...

Page 18: ...the post when turning the upper portion clockwise See assembling instruction step 9 4 Not able to tighten post enough to hide the red mark Have you extended the post long enough a Release pressure usi...

Page 19: ...and layout Does the pivoting track need to be more centred between the two standard posts See figure below Pivoting post may need to be moved so it is more centred between the two standard posts Follo...

Page 20: ...vel post accordingly Red mark on top of each post under the top plate is not visible The tracks are secured in place Press down on each track to make sure they are locked onto the posts Clean the insi...

Page 21: ...0 kg 440 lb Range Fixed Track 1900 3160 mm 74 124 1 2 Pivot track 1900 3160 mm 74 124 1 2 Post length 2130 2745 mm 84 108 Material Aluminium END OF LIFE DISPOSAL Package Corrugated cardboard recyclabl...

Page 22: ...Page intentionally left blank...

Page 23: ...6 87426222 E mail Italy promo arjo com MIDDLE EAST Arjo Middle East FZ LLC Office 908 9th Floor HQ Building North Tower Dubai Science Park Al Barsha South P O Box 11488 Dubai United Arab Emirates Dire...

Page 24: ...d health challenges With products and solutions that ensure ergonomic patient handling personal hygiene disinfection diagnostics and the effective prevention of pressure ulcers and venous thromboembol...

Reviews: