001.10610.EN rev
. 13
Room Assessment
7
Before installing the
EASYTRACK
prepare the installation as follows:
1) Survey the room where you plan to install the
EASYTRACK
System.
•
Make sure the room meets all the specifications
outlined in
“EASYTRACK System Requirements”
section.
2) Make sure the pick-up point will be inside the transfer
zone:
• Over a bed :
approximately 915 - 1020 mm (36 - 40”) from the head
of the bed.
• Over a tub:
approximately 400 - 460 mm (16 - 18”) from the end
of the tub.
• Allow 1000 mm (39”) of the open area for transfer
working space.
3) Determine the location for the three posts considering the
following, see Fig. 6:
• Distance from the wall, adjacent to the two standard
posts, to allow clearance for the pivot track overhang.
• Minimum distance of 1.9 m (75”) between the two
standard posts.
• Transfers should only take place between the
EASYTRACK
System’s posts.
• Centre the pivot post as much as possible between
both standard posts.
Fig. 5
400 - 460 mm
(16 - 18’’)
915 - 1020 mm
(36 - 40’’)
Bed
Bath
tub
Ø 1 m (39”)
Ø 1 m (39”)
Room Assessment
Fig. 6
NOTE:
Although it is not a requirement to
place the posts directly under a joist in the
ceiling, it is preferable. It does not matter which
direction the plates are facing, see Fig. 7
Fig. 7
Joists or
Trusses
EASYTRACK
POSTS
Sheet rock
Preferable
Joists or
Trusses
Acceptable
EASYTRACK POST
Sheet rock