7. Common Malfunction & Troubleshooting Solutions
(1)
The equipment can't be started?
A.
Make sure the power cord is plugged into a good power socket.
B.
Check whether the fuse in the back is loose or burnt out.
(2)
The equipment can be started, but the RF doesn't output energy, or the energy is weakened?
A.
Please check whether the handpiece and the plug connecting to the equipment body are closely
linked.
B.
The wire inside is loose or burnt out.
(3)
The equipment can be started, but there is an error message on the monitor?
A.
Take out the plug in the back, wait for about one minute, re-plug it, and restart the equipment.
B.
If the solutions mentioned above are not working, please contact the equipment dealer for assistance.
(4)
The Body Vacuum RF doesn't have suction, or its suction is weakened?
A.
Check whether the filter is installed correctly.
B.
Check whether the connecting wire of the handle is installed correctly.
C.
Check whether the connecting wire of the handle is damaged.
D.
There is foreign matter or oil dirt inside the air inlet of the main machine. Please check and clean it.
(5)
The equipment can be started, but the BIO doesn't have current output?
A.
Check whether the connecting wire of the handpiece is installed correctly.
B.
Check whether the energy intensity is set or the handpiece is turned on.
C.
Please check whether the products used are the adaptive ones specified by the
equipment (this handpiece is most suited to be used with liquid-state products).
(6)
The equipment can be started, but the Cold Hammer is not working?
A.
The cooling fan is damaged.
B.
The chipboard inside is loose or burnt out.
C.
If the situations mentioned above apply, the handpiece should be replaced.