p3
2. SPACE REQUIREMENTS
130
80
E 5 . . . . . . E 1
S2 S1
~
Dimensions:
without µconsole: 22.5 x 75 x 120 mm
with µconsole: 26.5 x 80 x130 mm
Housing:
self-extinguishing case of black UL 94VO ABS
Mounting in switchbox: latching on symmetrical DIN rail.
Rack version: consult.
Pug-off connectors for screwed connectings
(2.5mm², flexible or rigid).
Protection:
Case / terminals: IP 20
Operating temperature:
-5 to 50°C
Storage temperature:
-20 to 70°C
Upper connector
Lower connectors
26.5
programming
keys
3. CONNECTINGS
To allow the inserting of the µconsole and
ensure a sufficient thermic dissemination:
mount the instruments vertically (on horizontal
DIN rail) leaving a 10mm space between each.
x
22.5
x
22.5
x
26.5
x
26.5
10
8
6
A3
A1
A2
E3
E4
E5
E1
E2
PT100 3 wire
AC
~ ~
DC
- +
U p p e r c o n n e c t o r s
Lower connector
INPUT
ANALOG OUTPUT
S2 S1
mA
0-4/20 mA
active
Current
output
+
-
A3 A2 A1
Connecting
: In order to keep the safety the connecting must be
followed strictly, especially the
setting to the earth of terminal A2
in the group «supply».
SUPPLY