FastSCAN
49
10.3
Alignment Check (Scorpion Users Only)
The Alignment Check is a simple test to assess the alignment of the optics and is
independent of the tracking system.
Perform an Alignment Check as follows:
1. From the
FastSCAN
TM
software, select
Scanner/Alignment Check
.
2. Position a sheet of white paper (US-letter or A4) near the Transmitter.
3. Set the sensitivity control to position 1.
4. Gently place the Wand on to the center of the sheet of paper so that the laser
aperture faces downwards and the Wand lies across the width of the sheet.
5. With the trigger pulled fully in lift the Wand vertically off the sheet at a constant
rate. Note the arrangement of Wand, Transmitter and test paper as shown in
Figure 31.
6. Release the trigger once the laser line exceeds the width of the sheet.
7. Press OK.
Figure 31: Procedure for performing an Alignment Check.
The result of the Alignment Check is a number greater than zero. A lower number
means a better alignment. If the number is greater than 2.0 the Wand requires cali-
bration correction (see Section 10).