![AR 350AH1A Operating And Service Manual Download Page 41](http://html1.mh-extra.com/html/ar/350ah1a/350ah1a_operating-and-service-manual_2962316041.webp)
Model 350AH1A
Rev B
25
2.6.6.13
Output Measurement Display Mode
Command to set the format used to display output measurements.
Syntax:
DISP:x
Parameters:
Mode(x):
IVR
= display RMS Voltage and RMS Current
PWR
= display RMS Power
Response Format:
None (No query for this command)
Example:
To display the RMS output voltage and RMS output current, send the following
command:
DISP:IVR<LF>
To display the RMS output power, send the following command:
DISP:PWR<LF>
2.6.6.14
Faults
Query to find the faults that have occurred with the amplifier.
Syntax:
FSTA?
Parameters:
None
Response Format:
FSTA=<space>xxxx<LF>
Where:
xxxx
=
0000
to
001A
(Hexadecimal)
Dec
xxxx
Fault Text
Description
RF State After
Fault Reset
0
0000
N/A
No fault
N/A
1
0001
AC Interlock
AC Interlock
Off
4
0004
Interlock
Interlock
Off
5
0005
PS2
Power Supply 2
Off
22
0016
Thermal Fault
Thermal
Off
26
001A
ALC
ALC Fault
Off
Example:
To find out what faults have occurred, send the following query:
FSTA?<LF>
Response:
FSTA= 0004<LF>
(Interlock Fault)