Model 5000A225A
Rev B
25
2.4.2.6.3
Reset Faults
This will clear all faults, if possible.
Syntax:
RESET
Parameters:
None
Response Format: None (No query for this command)
Example:
To clear any faults, send the following command:
RESET<LF>
2.4.2.6.4
Mode Select
This command sets the ALC mode of the Model 5000A225A.
Syntax:
MODE:x
Parameters:
Mode(x):
MANUAL
= Set to Manual/CW mode
ALC<space>INT
= Set to ALC Internal mode
Response Format: None (No query for this command)
Example:
To set the ALC mode to Manual/CW mode, send the following command:
MODE:MANUAL<LF>
2.4.2.6.5
Level Adjust
This command sets the RF gain, detector gain, ALC threshold, and ALC Response Time of the Model
5000A225A.
Syntax:
LEVEL:xy
Parameters:
Parameter(x):
GAIN
= RF Gain
DET
= Detector Gain
THR
= ALC Threshold
RESP
= Response Time
Value(y):
For RF Gain, Detector Gain, and ALC Threshold:
0
= Minimum
100
= Maximum
For Response Time:
1
= Minimum
7
= Maximum